基于频率的母体心电图衰减在胎儿心电图分析中的应用。

Frequency-Based Maternal Electrocardiogram Attenuation for Fetal Electrocardiogram Analysis.

机构信息

Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Hospital, 111 Michigan Ave. NW, Washington, DC, 20010, USA.

Division of Cardiology, Children's National Hospital, Washington, DC, USA.

出版信息

Ann Biomed Eng. 2022 Jul;50(7):836-846. doi: 10.1007/s10439-022-02959-4. Epub 2022 Apr 11.

Abstract

Fetal electrocardiogram (ECG) waveform analysis along with cardiac time intervals (CTIs) measurements are critical for the management of high-risk pregnancies. Currently, there is no system that can consistently and accurately measure fetal ECG. In this work, we present a new automatic approach to attenuate the maternal ECG in the frequency domain and enhance it with measurable CTIs. First, the coherent components between the maternal ECG and abdominal ECG were identified and subtracted from the latter in the frequency domain. The residual was then converted into the time domain using the inverse Fourier transform to yield the fetal ECG. This process was improved by averaging multiple beats. Two fetal cardiologists, blinded to the method, assessed the quality of fetal ECG based on a grading system and measured the CTIs. We evaluated the fetal ECG quality of our method and time-based methods using one synthetic dataset, one human dataset available in the public domain, and 37 clinical datasets. Among the 37 datasets analyzed, the mean average (± standard deviation) grade was 3.49 ± 1.22 for our method vs. 2.64 ± 1.26 for adaptive interference cancellation (p-value < 0.001), thus showing the frequency-based fetal ECG extraction was the superior method, as assessed from our clinicians' perspectives. This method has the potential for use in clinical settings.

摘要

胎儿心电图(ECG)波形分析和心脏时间间隔(CTI)测量对于高危妊娠的管理至关重要。目前,没有能够始终如一地准确测量胎儿 ECG 的系统。在这项工作中,我们提出了一种新的自动方法,用于在频域中衰减母体 ECG 并增强可测量的 CTI。首先,在频域中识别母体 ECG 和腹部 ECG 之间的相干分量,并从后者中减去。然后,使用傅里叶逆变换将剩余部分转换为时域,以产生胎儿 ECG。通过平均多个心动周期来改进此过程。两位对该方法不知情的胎儿心脏病专家根据分级系统评估胎儿 ECG 的质量,并测量 CTI。我们使用一个合成数据集、一个公共领域中可用的人类数据集和 37 个临床数据集评估了我们的方法和基于时间的方法的胎儿 ECG 质量。在分析的 37 个数据集中,我们的方法的平均等级为 3.49 ± 1.22,而自适应干扰消除的平均等级为 2.64 ± 1.26(p 值<0.001),因此,从我们临床医生的角度来看,基于频域的胎儿 ECG 提取是更好的方法。该方法具有在临床环境中使用的潜力。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/fe3d/9148873/bffb760de2ee/10439_2022_2959_Fig1_HTML.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验