Scand J Clin Lab Invest. 2022 May;82(3):238-245. doi: 10.1080/00365513.2022.2056857. Epub 2022 Apr 11.
This study aims to compare skin perfusion pressure measurements (SPP) at midfoot and below knee level performed with a novel laser Doppler flowmetry (LD) probe with a reference method based on photo plethysmography (PP). It includes 40 patients referred with known or suspected peripheral arterial disease. The SPP was performed with both devices. Blinded re-interpretation of the SPP measurements was carried out by three observers and a consensus quality score was provided for each measurement. SPP >40 mmHg was considered a clinically relevant cut-off. This study evaluates a total of 48 paired measurements of the midfoot and 54 below knee. The two methods agreed in overall diagnostic classification in 80 of 102 measurements (78%) with both methods showing SPP ≥40 mmHg in 21 cases, and both methods showing SPP <40 mmHg in 59 cases. Of the 22 participants with disagreement (22%) in overall classification, 15 had measurements within the range of 30-50 mmHg, and 7 with a clinically relevant disagreement with one device showing SPP <30 mmHg and the other ≥40 mmHg. Analysis of inter observer variation for the LD readings showed an intraclass correlation coefficient of 0.880 (95% CI: 0.807 to 0.929, p- value <0.05) at midfoot, and 0.933 (95% CI: 0.894 to 0.959, -value <0.05) at below knee level. The novel probe based on LD showed good correlation with PP in absolute pressures, sufficient agreement in overall disease classification as well as good to excellent reliability in terms of inter observer variation.
本研究旨在比较使用新型激光多普勒流量仪(LD)探头和基于光体积描记法(PP)的参考方法在中足和膝下水平测量的皮肤灌注压(SPP)。研究纳入了 40 名已知或疑似外周动脉疾病的患者。使用两种设备进行 SPP 测量。由三位观察者对 SPP 测量结果进行盲法重新解释,并为每项测量提供共识质量评分。SPP>40mmHg 被认为是具有临床意义的截断值。本研究共评估了中足的 48 对和膝下的 54 对测量值。两种方法在 102 次测量中有 80 次(78%)的整体诊断分类一致,两种方法均显示 SPP≥40mmHg 的有 21 例,两种方法均显示 SPP<40mmHg 的有 59 例。在整体分类存在分歧的 22 名参与者(22%)中,有 15 名参与者的测量值在 30-50mmHg 范围内,有 7 名参与者存在具有临床意义的分歧,一种设备显示 SPP<30mmHg,另一种设备显示 SPP≥40mmHg。LD 读数的观察者间变异分析显示,中足的组内相关系数为 0.880(95%CI:0.807 至 0.929,p 值<0.05),膝下的组内相关系数为 0.933(95%CI:0.894 至 0.959,p 值<0.05)。基于 LD 的新型探头与 PP 在绝对压力方面具有良好的相关性,在整体疾病分类方面具有足够的一致性,在观察者间变异方面具有良好到极好的可靠性。