Suppr超能文献

钛合金股骨部件的早期失效:溶骨性和肉芽肿性改变的定量放射学分析

Early failure of titanium alloy femoral components: a quantitative radiological analysis of osteolytic and granulomatous change.

作者信息

Unwin A J, Stiles P J

机构信息

Department of Orthopaedics, Royal Surrey County Hospital, Guildford, UK.

出版信息

J R Soc Med. 1993 Aug;86(8):460-3. doi: 10.1177/014107689308600812.

Abstract

Shot blasted titanium alloy femoral components with a coarse finish were withdrawn from use in 1987 following reports of early component failure. We reviewed all coarse titanium femoral stems used in primary total hip arthroplasty in our institution between 1985 and 1987 and compared them to a similar stem made of vitallium which was also used over the same period. We found that, at an average of 5.2 years following implantation, over two-thirds of titanium components were associated with endosteal lysis. In many of the cases the early lytic changes were asymptomatic with significant bone loss occurring prior to presentation. We discuss our findings in relation to revision surgery. We advocate that all those with coarse stem titanium components should be positively reviewed to detect early lytic change before extensive bone destruction occurs.

摘要

1987年,有报道称表面粗糙的喷丸处理钛合金股骨部件出现早期部件失效,随后该部件停止使用。我们回顾了1985年至1987年间在我们机构用于初次全髋关节置换术的所有表面粗糙的钛股骨柄,并将它们与同期使用的一种类似的钴铬钼合金柄进行比较。我们发现,植入后平均5.2年,超过三分之二的钛部件出现骨内膜溶解。在许多病例中,早期的溶解变化是无症状的,在出现症状之前就已经发生了显著的骨质流失。我们讨论了与翻修手术相关的研究结果。我们主张,应对所有使用表面粗糙钛柄部件的患者进行积极检查,以便在广泛的骨质破坏发生之前检测到早期溶解变化。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/c3d8/1294051/ef52e8b183d7/jrsocmed00095-0031-a.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验