在用于放射治疗的 PET-CT-MR 三联体解决方案中定位不确定性的特性描述。

Characterization of positioning uncertainties in PET-CT-MR trimodality solutions for radiotherapy.

机构信息

QuantIF-LITIS EA4108, University of Rouen Normandie, Rouen, France.

GE Healthcare, Buc, France.

出版信息

J Appl Clin Med Phys. 2022 Jul;23(7):e13617. doi: 10.1002/acm2.13617. Epub 2022 Apr 28.

Abstract

The purpose of this study was to evaluate the positioning uncertainties of two PET/CT-MR imaging setups, C1 and C2. Because the PET/CT data were acquired on the same hybrid device with automatic image registration, experiments were conducted using CT-MRI data. In C1, a transfer table was used, which allowed the patient to move from one imager to another while maintaining the same position. In C2, the patient stood up and was positioned in the same radiotherapy treatment position on each imager. The two setups provided a set of PET/CT and MR images. The accuracy of the registration software was evaluated on the CT-MRI data of one patient using known translations and rotations of MRI data. The uncertainties on the two setups were estimated using a phantom and a cohort of 30 patients. The accuracy of the positioning uncertainties was evaluated using descriptive statistics and a t-test to determine whether the mean shift significantly deviated from zero (p < 0.05) for each setup. The maximum registration errors were less than 0.97 mm and 0.6° for CT-MRI registration. On the phantom, the mean total uncertainties were less than 2.74 mm and 1.68° for C1 and 1.53 mm and 0.33° for C2. For C1, the t-test showed that the displacements along the z-axis did not significantly deviate from zero (p = 0.093). For C2, significant deviations from zero were present for anterior-posterior and superior-inferior displacements. The mean total uncertainties were less than 4 mm and 0.42° for C1 and less than 1.39 mm and 0.27° for C2 in the patients. Furthermore, the t-test showed significant deviations from zero for C1 on the anterior-posterior and roll sides. For C2, there was a significant deviation from zero for the left-right displacements.This study shows that transfer tables require careful evaluation before use in radiotherapy.

摘要

本研究旨在评估两种 PET/CT-MR 成像设备(C1 和 C2)的定位不确定性。由于 PET/CT 数据是在具有自动图像配准功能的同一混合设备上采集的,因此使用 CT-MRI 数据进行了实验。在 C1 中,使用了一个转移台,允许患者在保持相同位置的情况下从一台成像仪移动到另一台成像仪。在 C2 中,患者站起来并在每台成像仪上处于相同的放射治疗位置。这两种设备提供了一组 PET/CT 和 MR 图像。使用已知的 MRI 数据平移和旋转在一位患者的 CT-MRI 数据上评估了配准软件的准确性。使用一个体模和 30 名患者的队列估计了两个设备的不确定性。使用描述性统计和 t 检验评估了定位不确定性的准确性,以确定每个设备的平均偏移量是否显著偏离零(p<0.05)。CT-MRI 配准的最大注册误差小于 0.97mm 和 0.6°。在体模上,C1 的总平均不确定性小于 2.74mm 和 1.68°,C2 的总平均不确定性小于 1.53mm 和 0.33°。对于 C1,t 检验表明 z 轴方向的位移没有显著偏离零(p=0.093)。对于 C2,前-后和上-下方向的位移存在显著偏离零的情况。C1 的患者平均总不确定性小于 4mm 和 0.42°,C2 的患者平均总不确定性小于 1.39mm 和 0.27°。此外,t 检验表明 C1 在前后和滚转方向上存在显著偏离零的情况。对于 C2,左右方向的位移存在显著偏离零的情况。本研究表明,在将转移台用于放射治疗之前,需要对其进行仔细评估。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/00a6/9278679/2436f0eb30e5/ACM2-23-e13617-g005.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验