Pearlin R Benedicta, Livingstone Roshan Samuel, Jasper Anita, Keshava Shyam Kumar N, Sridhar Gibikote
Department of Radiology, Christian Medical College, Vellore, Tamil Nadu, India.
J Med Phys. 2022 Jan-Mar;47(1):73-78. doi: 10.4103/jmp.jmp_127_21. Epub 2022 Mar 31.
To evaluate the image quality of semi-anatomical chest radiographs acquired using low radiation doses from seven different flat-panel detector (FPD) systems.
Radiographs of a semi-anatomical chest phantom were acquired at 70 and 110 kVp using 7 different FPDs from 5 vendors. Radiation doses were measured using a dose-area-product (DAP) meter. To standardize measurements across all FPDs, DAP value of 51.05 μGym obtained at 70 kVp and 9.43 μGym at 110 kVp was used as reference in this study. Radiation doses were reduced by manually adjusting mAs for both tube potentials in all FPD systems to achieve acceptable image quality. Contrast-to-noise ratio, signal-to-noise ratio and figure of merit (FOM) in lung, heart, and diaphragm for all images were analyzed.
In comparison with set standard protocol, radiation dose reduction of 20%, 21%, 21.6%, 59.5%, 60.7%, 62.2%, and 67.6% with optimal image quality was observed in Prognosys Prorad, GE Definium 8000, Siemens Fusion, Fujifilm FGX, Fujifilm FGXR, Philips Digital Diagnost and Siemens Aristos at 70 kVp. At 110 kVp, dose reduction of 15.7% and 34.8% was possible only for Philips Digital Diagnost and Siemens Aristos. FOM was high at 110 kVp even when radiation doses were reduced by a factor 2 when compared to 70 kVp in all digital radiography systems.
This study demonstrates the feasibility of using a semi-anatomical chest phantom in the optimization of radiation dose and image quality. The FOM was a good indicator in assessing image quality between different detectors.