将IDEXX SediVue Dx分析仪用于量化猫和狗尿液沉渣中红细胞(RBC)和白细胞(WBC)指标的准确性与手工显微镜评估的比较。

Accuracy of the IDEXX SediVue Dx analyzer for quantifying RBC and WBC indices in the urine sediments of cats and dogs compared with manual microscopic evaluations.

作者信息

Blanco Anthony E, Heseltine Johanna C, Hernandez Annalisa M, Bilbrough Graham E A, DeNicola Dennis B, Myrick Celine, Edwards Suzanne, Hammond Jeremy M, Myers Alexandra N, Nabity Mary B

机构信息

College of Veterinary Medicine and Biomedical Sciences, Texas A&M University, College Station, Texas, USA.

IDEXX Laboratories, Inc, Westbrook, Maine, USA.

出版信息

Vet Clin Pathol. 2022 Dec;51(4):470-479. doi: 10.1111/vcp.13141. Epub 2022 May 20.

Abstract

BACKGROUND

The IDEXX SediVue Dx (SediVue) is an automated, in-clinic urine sediment analyzer for veterinary patients. The bias between the results from manual microscopy and the SediVue is currently unknown.

OBJECTIVES

To assess the diagnostic accuracy of the SediVue, we aimed to determine the bias between the SediVue (index test) and manual microscopy (reference standard) for the quantification of RBCs and WBCs in urine.

METHODS

Urine remnant samples were collected from cats and dogs that contained RBCs (n = 462) and WBCs (n = 510). Retrospective analysis of results from urine sediment examinations using both manual microscopy (using a KOVA and DeciSlide system) and the SediVue (1.0.1.3) was performed. Bias was determined with Bland-Altman plots. SediVue-captured images from high-bias samples were reviewed, and biases were compared.

RESULTS

The median bias for semi-quantitative RBC and WBC counts was determined for RBC and WBC counts. The cutoffs were RBC ≤ 5/HPF, 0.3; RBC 5.1-10/HPF, 10.1; RBC 10.1-20/HPF, 10.6; and RBC > 20/HPF, 28.93; WBC ≤ 5/HPF, 0.1; WBC 5.1-10/HPF, 2.2; WBC 10.1-20/HPF, 9.4; and WBC > 20/HPF, 26.6. High bias between the methods was identified in 98 samples (21.0%) with RBCs and 77 samples (15.7%) with WBCs. Reviewer-based enumeration of the SediVue-captured images decreased the percentage of samples with high bias to 17.3% for RBCs and to 11.4% for WBCs.

CONCLUSIONS

Bias in the RBC and WBC counts between manual microscopy and the SediVue was unlikely to impact clinical interpretations in a majority of cases. Although reviewer enumeration of SediVue-captured images reduced observed bias, inherent differences between methodologies appeared to have a larger impact on the bias.

摘要

背景

IDEXX SediVue Dx(SediVue)是一款用于兽医患者的自动化临床尿液沉渣分析仪。目前尚不清楚手动显微镜检查结果与SediVue之间的偏差情况。

目的

为评估SediVue的诊断准确性,我们旨在确定SediVue(指标检测)与手动显微镜检查(参考标准)在尿液中红细胞(RBC)和白细胞(WBC)定量方面的偏差。

方法

收集含有RBC(n = 462)和WBC(n = 510)的猫和狗的尿液残留样本。对使用手动显微镜检查(使用KOVA和DeciSlide系统)和SediVue(1.0.1.3)进行的尿液沉渣检查结果进行回顾性分析。通过Bland-Altman图确定偏差。对来自高偏差样本的SediVue捕获图像进行审查,并比较偏差。

结果

确定了RBC和WBC计数的半定量RBC和WBC计数的中位数偏差。截断值为:RBC≤5/HPF,偏差为0.3;RBC 5.1 - 10/HPF,偏差为10.1;RBC 10.1 - 20/HPF,偏差为10.6;RBC>20/HPF,偏差为28.93;WBC≤5/HPF,偏差为0.1;WBC 5.1 - 10/HPF,偏差为2.2;WBC 10.1 - 20/HPF,偏差为9.4;WBC>20/HPF,偏差为26.6。在98份含RBC样本(21.0%)和77份含WBC样本(15.7%)中发现两种方法之间存在高偏差。基于审阅者对SediVue捕获图像的计数,使含RBC样本的高偏差百分比降至17.3%,含WBC样本的高偏差百分比降至11.4%。

结论

在大多数情况下,手动显微镜检查与SediVue之间RBC和WBC计数的偏差不太可能影响临床解读。尽管审阅者对SediVue捕获图像的计数减少了观察到的偏差,但方法学之间的固有差异似乎对偏差有更大影响。