Johnson V, Israel C, Schmitt J
Med. Klinik I, Abteilung für Kardiologie, UKGM Gießen, Universitätsklinik Gießen, Klinikstr. 33, 35392, Gießen, Deutschland.
Klinik für Innere Medizin, Kardiologie, Diabetologie und Nephrologie, Ev. Krankenhaus Bielefeld, Bielefeld, Deutschland.
Herzschrittmacherther Elektrophysiol. 2022 Jun;33(2):247-254. doi: 10.1007/s00399-022-00864-w. Epub 2022 May 23.
Programming of implantable cardiac devices, especially dual-chamber pacemakers, can be challenging in daily clinical practice. Precise knowledge of programmable parameters is important; furthermore, one should also be familiar with the specific algorithms of each manufacturer. During programming, the patient's individual requirements should be taken into account, but out-of-the-box programming should be avoided. Another important goal of programming should be to stimulate as much as needed but as little as possible to provide the patient good exercise capacity while not being aware of the pacing. Manufacturers' algorithms can help reach these aims but need to be understood and-in case of inappropriate behavior-to be deactivated.
在日常临床实践中,植入式心脏设备的编程,尤其是双腔起搏器的编程可能具有挑战性。精确了解可编程参数很重要;此外,还应熟悉每个制造商的特定算法。在编程过程中,应考虑患者的个体需求,但应避免开箱即用的编程方式。编程的另一个重要目标应该是按需尽可能多地刺激,但尽可能少地刺激,以便在患者未察觉到起搏的情况下为其提供良好的运动能力。制造商的算法有助于实现这些目标,但需要理解这些算法,并且在出现不当行为时将其停用。