Siegmund J B, Wilson J H, Lattner S E, Granneman K, Johnson R
Department of Cardiology, Good Samaritan Hospital, Cincinnati, Ohio 45220,USA.
Pacing Clin Electrophysiol. 1996 Jan;19(1):90-4. doi: 10.1111/j.1540-8159.1996.tb04795.x.
This study compares three different pacing system analyzers, which measure impedance at different points during an impulse, with measurements telemetered from implanted pacemakers from four different manufacturers. Measurements were obtained at the time of implantation in a group of 103 patients. The measurements obtained by these different methods differ significantly from each other; the later during the impulse the impedance is measured, the higher in general its value. Interpretation of impedance values should take into account the technique of measurement.