Suppr超能文献

助听器起始反应与稳态反应的比较:对听性脑干反应在助听器选配中应用的启示

Comparison of onset and steady-state responses of hearing aids: implications for use of the auditory brainstem response in the selection of hearing aids.

作者信息

Gorga M P, Beauchaine K A, Reiland J K

出版信息

J Speech Hear Res. 1987 Mar;30(1):130-6. doi: 10.1044/jshr.3001.130.

Abstract

Input-output (I/O) functions of hearing aids were measured in response to a 2000-Hz tone burst, having 0.5 ms rise/fall time and 10 ms duration. I/O functions, measured with a hearing-aid analyzer, served as reference conditions. Hearing-aid outputs at onset and during the steady-state portion of the waveform differed; these differences often depended upon stimulus rate. The relation between onset and steady-state estimates of output were not always predictable from hearing-aid attack and release times. These findings indicate that the steady-state output limitation characteristics of hearing aids cannot be estimated from their onset responses. In turn, this suggests that ABR measurements may not provide accurate estimates of the compressive characteristics of hearing aids.

摘要

针对一个2000赫兹的短纯音(上升/下降时间为0.5毫秒,持续时间为10毫秒),测量了助听器的输入-输出(I/O)功能。使用助听器分析仪测量的I/O功能作为参考条件。波形起始点和稳态部分的助听器输出不同;这些差异通常取决于刺激速率。输出的起始点和稳态估计值之间的关系并不总是能从助听器的启动和释放时间预测出来。这些发现表明,无法从助听器的起始反应来估计其稳态输出限制特性。反过来,这表明听觉脑干反应测量可能无法准确估计助听器的压缩特性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验