脑脊液和血清游离轻链κ检测的实验室间一致性。

Inter-Laboratory Concordance of Cerebrospinal Fluid and Serum Kappa Free Light Chain Measurements.

机构信息

Department of Laboratory Medicine, Azienda Ospedaliero-Universitaria and Azienda Unità Sanitaria Locale, 41126 Modena, Italy.

Department of Biomedical, Metabolic and Neurosciences, University of Modena and Reggio Emilia, 41126 Modena, Italy.

出版信息

Biomolecules. 2022 May 7;12(5):677. doi: 10.3390/biom12050677.

Abstract

The kappa index (K-Index), calculated by dividing the cerebrospinal fluid (CSF)/serum kappa free light chain (KFLC) ratio by the CSF/serum albumin ratio, is gaining increasing interest as a marker of intrathecal immunoglobulin synthesis. However, data on inter-laboratory agreement of these measures is lacking. The aim was to assess the concordance of CSF and serum KFLC measurements, and of K-index values, across different laboratories. KFLC and albumin of 15 paired CSF and serum samples were analyzed by eight participating laboratories. Four centers used Binding Site instruments and assays (B), three used Siemens instruments and assays (S), and one center used a Siemens instrument with a Binding Site assay (mixed). Absolute individual agreement was calculated using a two-way mixed effects intraclass correlation coefficient (ICC). Cohen's kappa coefficient (k) was used to measure agreement on positive (≥5.8) K-index values. There was an excellent agreement in CSF KFLC measurements across all laboratories (ICC (95% confidence interval): 0.93 (0.87-0.97)) and of serum KFLC across B and S laboratories (ICC: 0.91 (0.73-0.97)), while ICC decreased (to 0.81 (0.53-0.93)) when including the mixed laboratory in the analysis. Concordance for a positive K-Index was substantial across all laboratories (k = 0.77) and within S laboratories (k = 0.71), and very good (k = 0.89) within B laboratories, meaning that patients rarely get discordant results on K-index positivity notwithstanding the testing in different laboratories and the use of different platforms/assays.

摘要

脑脊液(CSF)/血清游离轻链κ(KFLC)比值除以 CSF/血清白蛋白比值的κ 指数(K-Index)作为鞘内免疫球蛋白合成的标志物越来越受到关注。然而,关于这些指标的实验室间一致性的数据尚缺乏。本研究旨在评估不同实验室间 CSF 和血清 KFLC 测量值以及 K-Index 值的一致性。8 家参与实验室分析了 15 对 CSF 和血清样本的 KFLC 和白蛋白。4 个中心使用 Binding Site 仪器和检测方法(B),3 个中心使用西门子仪器和检测方法(S),1 个中心使用西门子仪器和 Binding Site 检测方法(混合)。使用双向混合效应组内相关系数(ICC)计算绝对个体一致性。使用 Cohen 的 kappa 系数(k)来衡量 K-Index 值为阳性(≥5.8)的一致性。所有实验室的 CSF KFLC 测量值具有极好的一致性(ICC(95%置信区间):0.93(0.87-0.97)),B 和 S 实验室的血清 KFLC 具有极好的一致性(ICC:0.91(0.73-0.97)),而包括混合实验室时 ICC 降低(至 0.81(0.53-0.93))。所有实验室的阳性 K-Index 一致性很高(k = 0.77),S 实验室的一致性较高(k = 0.71),B 实验室的一致性非常好(k = 0.89),这意味着尽管在不同的实验室进行检测并且使用不同的平台/检测方法,患者很少会出现 K-Index 阳性的不一致结果。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/42ad/9138559/dfb078f8e844/biomolecules-12-00677-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验