Osawa Yosuke, Dohi Tetsuji
Micro System Laboratory, Chuo University, 1-13-27, Kasuga, Bunkyo-ku, Tokyo 112-8551, Japan.
Micromachines (Basel). 2022 Apr 27;13(5):679. doi: 10.3390/mi13050679.
In this paper, we report a new discriminant index for evaluating the wearing condition of a blood pressure pulse wave measurement device with small variability. The prototype device consists of MEMS 3-axis force sensors, a 3-axis push-in adjustment mechanism, a fixed jig, and a signal processing board. The arterial tonometry method is used by this device to measure blood pressure pulse waves. However, to accurately measure the blood pressure pulse wave, the device must be appropriately worn. The evaluation of the wearing condition using the amplitude of the blood pressure pulse wave was complicated by the wearing process due to the large variability. Therefore, we propose the ratio of the blood pressure pulse wave component to the noise component as an index that can evaluate the wear condition of the device. The correlation coefficient between the discriminant index and the amplitude of the blood pressure pulse wave had a correlation coefficient of 0.879. Furthermore, the discriminant index reduced the coefficient of variation by 12.8%. Therefore, it was suggested that the discriminant index is an index that can evaluate the pressing amount, one of the wearing conditions of the blood pressure pulse wave measurement device with a small influence of variation.
在本文中,我们报告了一种新的判别指标,用于评估具有小变异性的血压脉搏波测量装置的佩戴状况。该原型装置由MEMS三轴力传感器、三轴推入式调节机构、固定夹具和信号处理板组成。该装置采用动脉张力测量法来测量血压脉搏波。然而,为了准确测量血压脉搏波,该装置必须正确佩戴。由于变异性大,在佩戴过程中利用血压脉搏波的幅度来评估佩戴状况变得复杂。因此,我们提出将血压脉搏波分量与噪声分量的比值作为一种能够评估装置佩戴状况的指标。判别指标与血压脉搏波幅度之间具有0.879的相关系数。此外,判别指标使变异系数降低了12.8%。因此,有人认为该判别指标是一种能够评估按压量的指标,按压量是血压脉搏波测量装置佩戴状况之一,且受变异性影响较小。