Royal Adelaide Hospital, Adelaide, Australia; Centre for Heart Rhythm Disorders, the University of Adelaide, Adelaide, Australia.
Centre for Heart Rhythm Disorders, the University of Adelaide, Adelaide, Australia.
J Electrocardiol. 2022 Jul-Aug;73:42-48. doi: 10.1016/j.jelectrocard.2022.05.008. Epub 2022 May 21.
Little data exists on electrogram sensing in current generation of miniaturized insertable cardiac monitors (ICMs).
To compare the sensing capability of ICM with different vector length: Medtronic Reveal LINQ (~40 mm) vs. Biotronik Biomonitor III (BM-III, ~70 mm).
De-identified remote monitoring transmissions from n = 40 patients with BM-III were compared with n = 80 gender and body mass index (BMI)-matched patients with Reveal LINQ. Digital measurement of P- and R-wave amplitude from calibrated ICM electrograms was undertaken by 3 investigators independently. Further, we evaluated the impact of BMI and gender on P-wave visibility.
Patients in both groups were well matched for gender and BMI (53% male, mean BMI 26.7 kg/m2, both p = NS). Median P- and R-wave amplitude were 97% & 56% larger in the BM-III vs. LINQ [0.065 (IQR 0.039-0.10) vs. 0.033 (IQR 0.022-0.050) mV, p < .0001; & 0.78 (IQR 0.52-1.10) vs. 0.50 (IQR 0.41-0.89) mV, p = .012 respectively). The P/R-wave ratio was 36% greater with the BM-III (p < .001). The 25th percentile of P-wave amplitude for all 120 patients was .026 mV. Logistic regression analysis showed BM-III was more likely than LINQ to have P-wave amplitude ≥.026 mV (OR 7.47, 95%CI 1.965-29.42, p = .003), and increasing BMI was negatively associated with P-wave amplitude ≥.026 mV (OR 0.84, 95%CI 0.75-0.95, p = .004). However, gender was not significantly associated with P-wave amplitude ≥.026 mV (p = .37).
The longer ICM sensing vector of BM-III yielded larger overall P- and R- wave amplitude than LINQ. Both longer sensing vector and lower BMI were independently associated with greater P-wave visibility.
关于当前代小型可植入式心脏监测器(ICM)中的电描记法感应,数据很少。
比较不同向量长度的 ICM 的感应能力:美敦力 Reveal LINQ(40mm)与百多力 Biomonitor III(BM-III,70mm)。
对 n=40 名 BM-III 患者的远程监测传输数据进行去识别,并与 n=80 名性别和体重指数(BMI)匹配的 Reveal LINQ 患者进行比较。由 3 名研究人员独立对经校准的 ICM 电描记法的 P 波和 R 波振幅进行数字测量。此外,我们评估了 BMI 和性别对 P 波可见度的影响。
两组患者的性别和 BMI 均匹配良好(53%为男性,平均 BMI 为 26.7kg/m2,均为 p=NS)。与 LINQ 相比,BM-III 中的 P 波和 R 波振幅中位数分别大 97%和 56%[0.065(IQR 0.039-0.10)比 0.033(IQR 0.022-0.050)mV,p<0.0001;0.78(IQR 0.52-1.10)比 0.50(IQR 0.41-0.89)mV,p=0.012]。BM-III 的 P/R 波比值大 36%(p<0.001)。所有 120 名患者的 P 波振幅第 25 百分位数为.026mV。逻辑回归分析显示,与 LINQ 相比,BM-III 更有可能具有 P 波振幅≥.026mV(比值比 7.47,95%CI 1.965-29.42,p=0.003),而 BMI 增加与 P 波振幅≥.026mV 呈负相关(比值比 0.84,95%CI 0.75-0.95,p=0.004)。然而,性别与 P 波振幅≥.026mV 无显著相关性(p=0.37)。
与 LINQ 相比,BM-III 的较长 ICM 感应向量产生了更大的整体 P 波和 R 波振幅。较长的感应向量和较低的 BMI 都与 P 波可见度的增加独立相关。