全膝关节置换术中传统骨水泥固定的全聚乙烯胫骨部件的微动。迁移和诱导位移的X线立体摄影测量分析。

Micromotion of conventionally cemented all-polyethylene tibial components in total knee replacements. A roentgen stereophotogrammetric analysis of migration and inducible displacement.

作者信息

Ryd L, Lindstrand A, Rosenquist R, Selvik G

出版信息

Arch Orthop Trauma Surg (1978). 1987;106(2):82-8. doi: 10.1007/BF00435419.

Abstract

Micromotion of the tibial component in 27 knee arthroplasties for gonarthrosis, three of which were regarded as failures for reasons other than mechanical loosening, was studied using high-resolution roentgen stereophotogrammetric analysis (RSA). Migration over time was found in all cases, ranging from 0.2 to 2.1 mm. The majority of the cases showed migration during the first year only, while a minority migrated continuously during the period studied. The latter group showed more migration even after 1 year. All cases showed inducible displacement in response to external forces, ranging from 0.2 to 1.0 mm. The micromotion correlated with the age of the patient but not with other demographic parameters. Alignment or component position did not correlate with the micromotion. All cases but one had developed a radiolucent zone, and it was concluded that the micromotion occurred in the soft tissue interface between bone and cement. Such micromotion is a normal finding in cemented tibial components in knee arthroplasty.

摘要

使用高分辨率X线立体摄影测量分析(RSA)研究了27例膝关节骨关节炎置换术中胫骨部件的微动情况,其中3例因机械松动以外的原因被视为失败病例。所有病例均发现随时间的移位,范围为0.2至2.1毫米。大多数病例仅在第一年出现移位,而少数病例在研究期间持续移位。后一组在1年后仍有更多移位。所有病例对外力均表现出可诱导的位移,范围为0.2至1.0毫米。微动与患者年龄相关,但与其他人口统计学参数无关。对线或部件位置与微动无关。除1例病例外,所有病例均出现了透亮区,得出的结论是微动发生在骨与骨水泥之间的软组织界面。这种微动在膝关节置换术中骨水泥固定的胫骨部件中是正常现象。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验