Clarke W L, Santiago J V
Artif Organs. 1977 Nov;1(2):78-82. doi: 10.1111/j.1525-1594.1977.tb00982.x.
A new glucose sensor has been developed for the continuous extracorporeal monitoring of blood glucose concentrations in conjunction with an artificial pancreatic beta cell simulator. This sensor has a rapid ninety percent response time (less than 1 min), is stable for periods of up to 140 hours of accumulated use and can be calibrated without the use of a reference glucose analyzer. Its reliability has been confirmed in over 100 clinical studies. A comparison of this sensor to other glucose sensors designed for artificial pancreatic beta cell devices is made and the sensor is shown to be the most advanced glucose sensor presently available for extracorporeal continuous glucose monitoring in man.
一种新型葡萄糖传感器已被开发出来,用于与人工胰腺β细胞模拟器联合进行体外连续血糖浓度监测。该传感器具有快速的90%响应时间(不到1分钟),在累计使用长达140小时的时间段内保持稳定,并且无需使用参考葡萄糖分析仪即可进行校准。其可靠性已在100多项临床研究中得到证实。对该传感器与其他为人工胰腺β细胞设备设计的葡萄糖传感器进行了比较,结果表明该传感器是目前可用于人体体外连续血糖监测的最先进的葡萄糖传感器。