Linnér E, Rumberger E, Draeger J
Klin Monbl Augenheilkd. 1987 Jan;190(1):30-3.
A method of simultaneous measurement of intraocular pressure and volume displacement is reported. An electronic pressure sensor (diameter 3 mm) was positioned in the centre of the base of a transparent applanation body (diameter 9 mm). The corneal applanation area could be changed by different applanation forces and was measured photographically. In isolated human and porcine eyes, sensor signal and intraocularly measured pressure were related linearly and independently of the applanation area. In human eyes in situ, the rigidity coefficient decreased with increasing pressure. Adhesive forces and stiffness of the cornea obviously only falsify tonometry at the margin of the applanation area.
报告了一种同时测量眼压和体积位移的方法。一个电子压力传感器(直径3毫米)位于一个透明压平体(直径9毫米)底部的中心。角膜压平面积可通过不同的压平力改变,并通过摄影测量。在离体的人眼和猪眼中,传感器信号与眼内测量的压力呈线性关系,且与压平面积无关。在原位人眼中,刚度系数随压力增加而降低。角膜的粘附力和刚度显然仅在压平区域边缘对眼压测量产生干扰。