Fagni L, Hugon M, Folco A, Imbert G
Undersea Biomed Res. 1987 Mar;14(2):161-8.
A pressure chamber, designed for microelectrode recordings in isolated tissues or organs an capable of withstanding pressures up to 200 bar, is described. The versatility of the vessel allows a wide variety of experimental configurations and several types of studies. It features a complete access, easy visibility, interchangeable tissue bath and micromanipulator modules, as continuous perfusion and temperature control of the preparation under pressure. The ability to move micropipettes in 1-micron steps allows single unit recording on a variety of in vitro preparations at normal or elevated gas pressures. Successful physiologic tests, using the hippocampal slice preparation, are described. The results testify to the reliability of the system and to the usefulness of this in vitro model to study the gas pressure effects on isolated networks of mammalian CNS.
本文描述了一种压力室,其设计用于在离体组织或器官中进行微电极记录,能够承受高达200巴的压力。该容器的多功能性允许进行多种实验配置和多种类型的研究。它具有完全可进入性、易于观察、可互换的组织浴和微操纵器模块,以及在压力下对制剂进行连续灌注和温度控制。能够以1微米的步长移动微量移液器,允许在正常或升高的气压下对各种体外制剂进行单细胞记录。文中描述了使用海马切片制剂进行的成功生理测试。结果证明了该系统的可靠性以及这种体外模型在研究气压对哺乳动物中枢神经系统离体网络影响方面的有用性。