In this work, a new image guidance system and protocols for delivering image-guided radiotherapy (IGRT) on the Imaging and Medical Beamline (IMBL) at the ANSTO Australian Synchrotron are introduced. The image guidance methods used and the resulting accuracy of tumour alignment in in vivo experiments are often under-reported. Image guidance tasks are often complex, time-consuming and prone to errors. If unchecked, they may result in potential mis-treatments. We introduce SyncMRT, a software package that provides a simple, image guidance tool-kit for aligning samples to the synchrotron beam. We have demonstrated sub-millimetre alignment using SyncMRT and the small-animal irradiation platform (the DynamicMRT system) on the IMBL. SyncMRT has become the standard for carrying out IGRT treatments on the IMBL and has been used in all pre-clinical radiotherapy experiments since 2017. Further, we introduce two quality assurance (QA) protocols to synchrotron radiotherapy on the IMBL: the Winston-Lutz test and hidden target test. It is shown that the presented QA tests are appropriate for picking up geometrical setup errors and assessing the end-to-end accuracy of the image guidance process. Together, these tools make image guidance easier and provide a mechanism for reporting the geometric accuracy of synchrotron-based IGRT treatments. Importantly, this work is scalable to other delivery systems, and is in continual development to support the upcoming veterinary radiotherapy trials on the IMBL.
在这项工作中,我们介绍了一种新的图像引导系统和协议,用于在澳大利亚同步加速器 ANSTO 的成像和医疗射束线 (IMBL) 上提供图像引导放射治疗 (IGRT)。在体内实验中,通常很少报告所使用的图像引导方法和肿瘤对准的精度。图像引导任务通常很复杂、耗时且容易出错。如果不加以检查,它们可能导致潜在的治疗失误。我们引入了 SyncMRT,这是一个软件包,提供了一个简单的图像引导工具包,用于将样本与同步加速器光束对准。我们已经在 IMBL 上使用 SyncMRT 和小动物辐照平台(DynamicMRT 系统)演示了亚毫米对准。自 2017 年以来,SyncMRT 已成为在 IMBL 上进行 IGRT 治疗的标准,并已用于所有临床前放射治疗实验中。此外,我们还介绍了两种用于 IMBL 上同步放射治疗的质量保证 (QA) 协议:Winston-Lutz 测试和隐藏目标测试。结果表明,所提出的 QA 测试适用于检测几何安装误差,并评估图像引导过程的端到端精度。这些工具一起使图像引导更加容易,并为报告基于同步加速器的 IGRT 治疗的几何精度提供了一种机制。重要的是,这项工作具有可扩展性,可以支持即将在 IMBL 上进行的兽医放射治疗试验,并在不断发展,以支持即将进行的兽医放射治疗试验。