Acta Orthop. 2022 Jul 14;93:658-664. doi: 10.2340/17453674.2022.3976.
BoneMaster (BM) is a thin electrochemically applied hydroxyapatite (HA) implant coating marketed with expectations of improved osseointegration properties but less polyethylene (PE) wear. We compared the midterm cup migration and PE wear of cementless porous-coated hemispherical cups with and without BM.
In this patient-blinded, randomized controlled trial, 53 patients with a mean age of 64 years (55-75) received total hip arthroplasty with a porous-coated (P) or porous and BoneMaster (PBM) coated Exceed cup and ArCom E1 infused PE. Patients were followed with RSA, Hip Osteoarthritis Outcome Score (HOOS), and Euro-Qol-5-3L (EQ-5D) at 3 and 6 months, and 1-, 2-, and 5-year follow-up.
At 5-year follow-up, total translation and maximum total point motion was 0.28 mm (95% CI 0.08; 0.47) and 0.52 mm (CI 0.12; 0.93) higher in the PBM group than in the P group. PE wear was comparable between PBM and P cups, and 2D wear rate from 1-year follow-up to last follow-up was 0.03 mm (CI 0.02-0.03). The 5-year anterior translation was 0.05 mm (CI -0.10 to 0.21) in the normal BMD group and 0.40 mm (CI 0.22-0.57) in the osteopenia group.
At 5-year follow-up, Exceed cups in the PBM group migrated more than in the P group but the PE wear rate was low and similar. This study does not indicate any advantage of additional BoneMaster coating compared with porous coating alone on cementless hemispherical cups with regards to migration, polyethylene wear, and clinical outcomes.
BoneMaster(BM)是一种薄的电化学应用的羟基磷灰石(HA)植入涂层,其具有改善骨整合特性的预期,但聚乙烯(PE)磨损较少。我们比较了带和不带 BoneMaster(BM)的非骨水泥多孔涂层半球形杯的中期杯迁移和 PE 磨损。
在这项患者盲、随机对照试验中,53 名平均年龄为 64 岁(55-75 岁)的患者接受了全髋关节置换术,使用多孔涂层(P)或多孔和 BoneMaster(PBM)涂层的 Exceed 杯和 ArCom E1 输注的 PE。患者在 3 个月、6 个月、1 年、2 年和 5 年随访时接受 RSA、髋关节骨关节炎结局评分(HOOS)和欧洲五维健康量表(EQ-5D)评估。
在 5 年随访时,PBM 组的总平移和最大总点运动分别为 0.28 毫米(95%置信区间 0.08;0.47)和 0.52 毫米(CI 0.12;0.93)高于 P 组。PBM 和 P 杯的 PE 磨损相当,从 1 年随访到最后随访的 2D 磨损率为 0.03 毫米(CI 0.02-0.03)。正常骨密度组的 5 年前方平移为 0.05 毫米(CI -0.10 至 0.21),骨质疏松组为 0.40 毫米(CI 0.22-0.57)。
在 5 年随访时,PBM 组的 Exceed 杯迁移量大于 P 组,但 PE 磨损率较低且相似。本研究表明,与单独多孔涂层相比,在非骨水泥半球形杯上添加 BoneMaster 涂层在迁移、聚乙烯磨损和临床结果方面没有任何优势。