De Jonge N, Fillié Y E, Deelder A M
J Immunol Methods. 1987 May 20;99(2):195-7. doi: 10.1016/0022-1759(87)90127-x.
In our laboratory serum components interfering in the immunoassay of the schistosome proteoglycan circulating anodic antigen (CAA) in serum necessitated us to develop a simple technique by which the non-specific reaction of negative control sera could be prevented. Trichloroacetic acid was added to serum samples to precipitate interfering (glyco-)proteins. After centrifugation, the supernatant was neutralized and used directly either in the ELISA or in the indirect haemagglutination. This method gave satisfactory results, i.e., negative control sera did no longer give false positive reactions, while the titre of the positive controls remained unaffected. This method could also be successfully applied for the pretreatment of urine samples.
在我们实验室中,血清成分会干扰血清中血吸虫蛋白聚糖循环阳极抗原(CAA)的免疫测定,因此我们需要开发一种简单的技术,以防止阴性对照血清出现非特异性反应。向血清样本中加入三氯乙酸,沉淀干扰性(糖)蛋白。离心后,将上清液中和,直接用于酶联免疫吸附测定(ELISA)或间接血凝试验。该方法取得了满意的结果,即阴性对照血清不再出现假阳性反应,而阳性对照的滴度不受影响。该方法也可成功应用于尿液样本的预处理。