Liu Yu, He Xiaowei, Zou Jingjing, Ouyang Xiuyun, Huang Chunrong, Yang Xiao, Wang Yu
South China University of Technology, Guangzhou, China.
National & Local United Engineering Lab of Rapid Diagnostic Test, Guangzhou Wondfo Biotech Co., Ltd., Guangzhou, China.
Front Bioeng Biotechnol. 2022 Jul 7;10:920972. doi: 10.3389/fbioe.2022.920972. eCollection 2022.
Early screening of cancer can effectively prolong survival time and reduce cancer mortality. However, the existing health-monitoring devices can only be carried out in professional laboratories, so large-scale early cancer screening in resource-limited settings is hardly achieved. To embrace the challenge, we developed a novel chemiluminescence immunoassay (CLIA) analyzer that does not require a professional operation. Then, it was applied to detect carbohydrate antigen 50 (CA50), a non-organ-specific tumor marker for screening various cancers. As a result, the analyzer exhibited excellent performance that the total assay time was only 15 min, and the detection limit reached 0.057 U ml. A coefficient of variance () less than 15% was well-controlled for both intra- and inter-assay precision, and the linear range was 0-500 U ml. More importantly, this analyzer can continuously detect 60 samples per hour without any professional paramedic. Finally, this analyzer has been applied to evaluate clinical samples and the detected results showed a good correlation with the clinical test results (correlation coefficient, 0.9958). These characteristics exactly meet large-scale and high-throughput early screening of cancer. Thus, this miniaturized analyzer for CA50 detection is promising to achieve early large-scale screening of cancer in the resource-limited grassroots community.
癌症的早期筛查能够有效延长生存期并降低癌症死亡率。然而,现有的健康监测设备只能在专业实验室中进行操作,因此在资源有限的环境中很难实现大规模的早期癌症筛查。为了应对这一挑战,我们开发了一种新型的无需专业操作的化学发光免疫分析(CLIA)分析仪。然后,将其应用于检测糖类抗原50(CA50),一种用于筛查各种癌症的非器官特异性肿瘤标志物。结果,该分析仪表现出优异的性能,总检测时间仅为15分钟,检测限达到0.057 U/ml。批内和批间精密度的变异系数(CV)均控制在15%以内,线性范围为0-500 U/ml。更重要的是,该分析仪无需任何专业医护人员即可每小时连续检测60个样本。最后,该分析仪已应用于评估临床样本,检测结果与临床检测结果显示出良好的相关性(相关系数为0.9958)。这些特性完全符合癌症的大规模和高通量早期筛查要求。因此,这种用于检测CA50的小型化分析仪有望在资源有限的基层社区实现癌症的早期大规模筛查。