Obesity (Silver Spring). 2022 Sep;30(9):1766-1777. doi: 10.1002/oby.23476. Epub 2022 Aug 3.
This study aimed to evaluate the agreement of commonly reported energy metabolism measurements obtained from two different whole-room indirect calorimeters (WRICs).
Nine healthy adult volunteers were evaluated over four separate 24-hour periods in a crossover design, twice in two different WRICs of different sizes, each operated according to the Room Indirect Calorimetry Operating and Reporting Standards published in 2020. The reproducibility of repeated measurements was quantified by the coefficient of variation (CV) and intraclass correlation coefficient (ICC).
The CVs between and within each WRIC for average 24-hour carbon dioxide production rate (VCO ) and oxygen consumption rate (VO ), 24-hour energy expenditure (EE), and respiratory exchange ratio ranged from 1.5% to 3.6%, whereas sleep EE ranged from 3.1% to 5.5%. CVs for macronutrient oxidation rates and spontaneous physical activity were higher, ranging from 9.2% to 38.1%. ICCs of VCO , VO , 24-hour EE, and energy expenditure at zero activity were >0.95, indicating excellent reproducibility, whereas ICCs for lipid oxidation, awake and fed thermogenesis, and sleep EE ranged from 0.55 to 0.92, indicating moderate to high reproducibility. ICCs for respiratory exchange ratio and carbohydrate and protein oxidation rates were lower (<0.70). Spontaneous physical activity showed high reproducibility within chambers (ICC = 0.88) but differed substantially between chambers (ICC = 0.23).
Cross-chamber reproducibility is high for common outcome measures assessed in the respiratory chamber. The results support efforts to promote standardization across WRICs to allow multicenter studies.
本研究旨在评估两种不同整体房间间接测热仪(WRIC)获得的常用能量代谢测量值的一致性。
采用交叉设计,9 名健康成年志愿者在 4 个不同的 24 小时周期内进行评估,两次在两种不同大小的不同 WRIC 中进行,每种 WRIC 均按照 2020 年发布的《房间间接测热操作和报告标准》进行操作。通过变异系数(CV)和组内相关系数(ICC)来量化重复测量的重现性。
每个 WRIC 之间和内部的平均 24 小时二氧化碳产生率(VCO )和氧气消耗率(VO )、24 小时能量消耗(EE)和呼吸交换率的 CV 范围为 1.5%至 3.6%,而睡眠 EE 的 CV 范围为 3.1%至 5.5%。用于宏量营养素氧化率和自发体力活动的 CV 更高,范围为 9.2%至 38.1%。VCO 、VO 、24 小时 EE 和零活动能量消耗的 ICC 均>0.95,表明具有极好的重现性,而脂质氧化、清醒和进食产热以及睡眠 EE 的 ICC 范围为 0.55 至 0.92,表明具有中等至高的重现性。呼吸交换率和碳水化合物及蛋白质氧化率的 ICC 较低(<0.70)。自发体力活动在腔室内具有高重现性(ICC=0.88),但在腔室之间差异很大(ICC=0.23)。
在呼吸室内评估的常见结果测量值中,跨腔室重现性很高。这些结果支持努力促进 WRIC 之间的标准化,以允许进行多中心研究。