两种新型全身间接测热仪测量能量消耗和宏量营养素氧化的验证。

Validation of energy expenditure and macronutrient oxidation measured by two new whole-room indirect calorimeters.

机构信息

Institute of Human Nutrition and Food Science, Christian-Albrechts University, Kiel, Germany.

D & S Consulting Services Inc., New York, New York, USA.

出版信息

Obesity (Silver Spring). 2022 Sep;30(9):1796-1805. doi: 10.1002/oby.23527. Epub 2022 Aug 4.

Abstract

OBJECTIVE

The aim of this study was to validate two new whole-room indirfect calorimeters according to Room Indirect Calorimetry Operating and Reporting Standards (RICORS 1.0).

METHODS

For technical validation, 16 propane combustion tests were performed to determine accuracy and precision of energy expenditure (EE) and ventilation rates of oxygen (VO ), carbon dioxide (VCO ), and respiratory exchange ratio (VCO /VO ). For biological validation, eight participants (mean [SD], age 24.1 [2.5] years; BMI 24.3 [3.1] kg/m ) underwent four 24-hour protocols under highly standardized conditions: (1) isocaloric sedentary, (2) fasting sedentary, (3) isocaloric active, and (4) fasting active. Reliability (coefficients of variation [CV]) and minimal detectable changes (MDC) were calculated for 24-hour EE, sleeping metabolic rate (SMR), physical activity energy expenditure (PAEE), thermic effect of food (TEF), and macronutrient oxidation rates.

RESULTS

Technical validation showed high reliability and recovery rates for VO (0.75% and 100.8%, respectively), VCO (0.49% and 100.6%), and EE (0.54% and 98.2%). Biological validation revealed CV and MDC for active conditions of 1.4% and 4.3% for 24-hour EE, 1.7% and 5.9% for SMR, and 30.2% and 38.4% for TEF, as well as 5.8% and 10.5% for PAEE, respectively. Mean CV and MDC for macronutrient oxidation rates were 9.9% and 22.9%, respectively.

CONCLUSIONS

The precision of 24-hour EE and SMR was high, whereas it was lower for PAEE and poor for TEF.

摘要

目的

本研究旨在根据间接量热法操作和报告标准(RICORS 1.0)验证两种新型整体房间间接量热计。

方法

为了进行技术验证,进行了 16 次丙烷燃烧测试,以确定能量消耗(EE)和氧气(VO )、二氧化碳(VCO )和呼吸交换率(VCO / VO )的通气率的准确性和精密度。为了进行生物学验证,在高度标准化的条件下,八名参与者(平均[标准差],年龄 24.1[2.5]岁;BMI 24.3[3.1]kg/m )进行了四个 24 小时的协议:(1)等热量静坐,(2)禁食静坐,(3)等热量活动和(4)禁食活动。计算了 24 小时 EE、睡眠代谢率(SMR)、体力活动能量消耗(PAEE)、食物热效应(TEF)和宏量营养素氧化率的可靠性(变异系数[CV])和最小可检测变化(MDC)。

结果

技术验证显示 VO(分别为 0.75%和 100.8%)、VCO(分别为 0.49%和 100.6%)和 EE(分别为 0.54%和 98.2%)的高可靠性和回收率。生物学验证显示活动条件下 24 小时 EE 的 CV 和 MDC 分别为 1.4%和 4.3%、SMR 的 1.7%和 5.9%、TEF 的 30.2%和 38.4%,以及 PAEE 的分别为 5.8%和 10.5%。宏量营养素氧化率的平均 CV 和 MDC 分别为 9.9%和 22.9%。

结论

24 小时 EE 和 SMR 的精度较高,而 PAEE 的精度较低,TEF 的精度较差。