Vaddi Chandra Mohan, Ramakrishna Paidakula, Ganesan Soundarya, Swamy Siddalinga, Anandan Hemnath, Babu Manas
Preeti Urology and Kidney Hospital, Hyderabad, India.
Cent European J Urol. 2022;75(2):199-204. doi: 10.5173/ceju.2022.0224. Epub 2022 May 20.
The aim of this article was to evaluate the durability of current generation fibreoptic flexible ureteroscopes (FURS), analyse factors that influence durability, identify reasons for premature damage of FURS and offer suggestions to expand its life span.
A total of 952 retrograde intrarenal surgeries (RIRS) done for upper tract calculi using 8 fibreoptic FURS, namely three Storz Flex X2, one Flex X2S, two Olympus URF- P6, two Olympus URF- P7, between March 2013 and December 2018, were reviewed retrospectively. All procedures were done by two consultants, in a single referral centre. Data relating to stone characteristics and flexible ureteroscopy procedure were retrieved from hospital database. The primary end point was damage of FURS requiring first repair.
The average stone burden was 14.59 ±3.37 mm (range 3-22 mm). Ureteral access sheath was used in 95.4% of cases, 36.7% of the cases were pre-stented. Mean ureteroscope durability was 119 procedures and mean ureteroscopy time was 71.99 hours of use before first repair. Prolonged laser usage time and increased usage of accessories had significant negative impact on longevity of FURS (p = 0.002, p = 0.036 respectively). Inadvertent laser fibre misfire and extreme torque caused premature FURS damage, at the end of 35 and 12 procedures respectively.
Current generation fibreoptic flexible ureteroscopes have a mean durability of 119 procedures. Anticipation of torque, knowledge of the common reasons for damage and meticulous handling is essential to maximise the durability of FURS.
本文旨在评估当代纤维光学软性输尿管镜(FURS)的耐用性,分析影响耐用性的因素,确定FURS过早损坏的原因,并提出延长其使用寿命的建议。
回顾性分析了2013年3月至2018年12月期间使用8台纤维光学FURS(3台史托斯Flex X2、1台Flex X2S、2台奥林巴斯URF-P6、2台奥林巴斯URF-P7)对上尿路结石进行的952例逆行性肾内手术(RIRS)。所有手术均由两名顾问在单一转诊中心完成。从医院数据库中检索与结石特征和软性输尿管镜手术相关的数据。主要终点是FURS首次维修所需的损坏情况。
平均结石负荷为14.59±3.37mm(范围3-22mm)。95.4%的病例使用了输尿管通路鞘,36.7%的病例进行了预先支架置入。输尿管镜的平均耐用性为119次手术,首次维修前输尿管镜的平均使用时间为71.99小时。激光使用时间延长和附件使用增加对FURS的使用寿命有显著负面影响(分别为p = 0.002,p = 0.036)。意外的激光纤维误发射和极端扭矩分别在35次和12次手术后导致FURS过早损坏。
当代纤维光学软性输尿管镜的平均耐用性为119次手术。预判扭矩、了解常见损坏原因并谨慎操作对于最大限度提高FURS的耐用性至关重要。