哪种软性输尿管镜(数字式与光纤式)能够轻松抵达困难的下极肾盏并具有更好的尖端偏转:在K-Box上的体外研究。一项PETRA评估。

Which Flexible Ureteroscopes (Digital vs. Fiber-Optic) Can Easily Reach the Difficult Lower Pole Calices and Have Better End-Tip Deflection: In Vitro Study on K-Box. A PETRA Evaluation.

作者信息

Dragos Laurian B, Somani Bhaskar K, Sener Emre T, Buttice Salvatore, Proietti Silvia, Ploumidis Achilles, Iacoboaie Catalin T, Doizi Steeve, Traxer Olivier

机构信息

1 Clinical Emergency County Hospital "Pius Branzeu" Timisoara, University of Medicine and Pharmacy "Victor Babes" Timisoara , Romania .

2 PETRA-Progress in Endourology, Technology and Research Association , Paris, France .

出版信息

J Endourol. 2017 Jul;31(7):630-637. doi: 10.1089/end.2017.0109. Epub 2017 Jun 13.

Abstract

BACKGROUND AND PURPOSE

Modern flexible ureteroscopes (fURSs) have good deflection, but despite this, approaching an acute angled calix can still be difficult. The goals of our in vitro study were to assess the ability of the available modern fURSs to effectively access the sharp angled calices and to compare the end-tip deflection of the various fiber-optic and digital fURSs.

MATERIALS AND METHODS

Using a bench-training model for FURS (K-Box, Porgès-Coloplast), we tried to access an acute angled calix with nine different fURSs (BOA vision, COBRA vision, R.Wolf; FLEX X, FLEX Xc, K.Storz; LithoVue, Boston Scientific; URF-P5, URF-P6, URF-V, URF-V2, Olympus). Passing the fURSs through a ureteral access sheath (ReTrace, Porgès-Coloplast), the maximum end-tip deflection for every fURS was measured with the tip extended out from the sheath at 1, 2, 3, and 4 cm. Two ranking methods were designed for scoring the fURSs, one based on total ranking points and the other on total degrees of deflection.

RESULTS

While all fiber-optic fURSs (except URF-P6) were able to access the sharp angled calix, none of the digital fURSs (except FLEX Xc) reached the difficult angled calix. Similarly, all fiber-optic fURSs had better end-tip deflection compared with the digital fURSs, except FLEX Xc, which was as deflectable as the fiber-optic fURSs. The fURSs showed an end-tip deflection (median difference of almost 21°) in favor of fiber-optic fURSs. Based on the scoring, the highest ranked fURS (best deflection) was FLEX X2 and the lowest ranked fURS (worst deflection) was URF-V2.

CONCLUSIONS

Digital fURSs were less effective in accessing the sharp angled calix and they had lesser end-tip deflection compared with the fiber-optic counterparts. When approaching a difficult lower pole calix, it might be better to use a fiber-optic fURS.

摘要

背景与目的

现代软性输尿管镜(fURS)具有良好的可弯曲性,但即便如此,接近锐角肾盏仍可能存在困难。我们体外研究的目的是评估现有现代fURS有效进入锐角肾盏的能力,并比较各种纤维光学和数字式fURS的尖端可弯曲度。

材料与方法

使用一种用于fURS的模拟训练模型(K-Box,保赫曼公司),我们尝试使用9种不同的fURS(BOA vision、COBRA vision、R.Wolf;FLEX X、FLEX Xc、卡尔史托斯公司;LithoVue,波士顿科学公司;URF-P5、URF-P6、URF-V、URF-V2,奥林巴斯公司)进入锐角肾盏。将fURS通过输尿管鞘(ReTrace,保赫曼公司),当尖端从鞘伸出1、2、3和4厘米时,测量每种fURS的最大尖端可弯曲度。设计了两种评分方法对fURS进行评分,一种基于总排名分数,另一种基于总可弯曲度。

结果

虽然所有纤维光学fURS(URF-P6除外)都能够进入锐角肾盏,但数字式fURS(FLEX Xc除外)均未到达困难角度的肾盏。同样,与数字式fURS相比,所有纤维光学fURS的尖端可弯曲度更好,FLEX Xc除外,其可弯曲度与纤维光学fURS相当。fURS显示出尖端可弯曲度(中位数差异近21°)有利于纤维光学fURS。基于评分,排名最高的fURS(最佳可弯曲度)是FLEX X2,排名最低的fURS(最差可弯曲度)是URF-V2。

结论

数字式fURS进入锐角肾盏的效果较差,与纤维光学fURS相比,其尖端可弯曲度较小。当接近困难的下极肾盏时,使用纤维光学fURS可能更好。