Rev Assoc Med Bras (1992). 2022 Jul;68(7):935-938. doi: 10.1590/1806-9282.20220093.
Finger splints are used as a treatment option in tendon, bone, and soft tissue injuries. For immobilization, custom-made splints and prefabricated finger splints are used. In splints made for immobilization, it is aimed to limit joint movement. The aim of our study is to reveal how much custom-made splints and prefabricated finger splints limit joint motion (flexion angle in proximal interphalangeal and distal interphalangeal joints).
Custom-made splints and prefabricated finger splints were applied to the second fingers of the dominant side in a total of 40 individuals, 20 women and 20 men, not having any health problems. Individuals were asked to flex and joint motion was measured with the iPhone compass application.
The mean distal interphalangeal joint angle values of the participants measured by prefabricated finger splints were found to be 24.27±8.29, and the mean distal interphalangeal joint angle values measured by custom-made splints was 0.52±1.50. There was a difference between the participants' distal interphalangeal joint angle values measured by prefabricated finger splints and custom-made splints (p<0.001). distal interphalangeal joint angle values measured with custom-made splints were significantly smaller than those measured with prefabricated finger splint. The mean of the participants' proximal interphalangeal joint angle values measured by prefabricated finger splints was 16.55±7.90, and the proximal interphalangeal joint angle values measured by custom-made splints was "0" for all participants. There was a difference between the participants' proximal interphalangeal joint angle values measured by prefabricated finger splints and custom-made splints (p<0.001). Distal interphalangeal joint angle values measured with custom-made splints were significantly smaller than those measured with prefabricated finger splints.
According to our study, custom-made splints can significantly reduce the flexion of the finger interphalangeal joints compared to prefabricated finger splints.
指夹板作为治疗肌腱、骨骼和软组织损伤的一种选择。为了固定,使用定制夹板和预制指夹板。在用于固定的夹板中,目的是限制关节运动。我们的研究旨在揭示定制夹板和预制指夹板限制关节运动的程度(近侧指间关节和远侧指间关节的弯曲角度)。
将定制夹板和预制指夹板应用于 40 名优势侧第二指的个体,其中 20 名女性,20 名男性,无任何健康问题。要求个体弯曲手指,并使用 iPhone 指南针应用测量关节运动。
通过预制指夹板测量的参与者的远侧指间关节角度平均值为 24.27±8.29,通过定制夹板测量的参与者的远侧指间关节角度平均值为 0.52±1.50。通过预制指夹板和定制夹板测量的参与者的远侧指间关节角度值之间存在差异(p<0.001)。通过定制夹板测量的远侧指间关节角度值明显小于通过预制指夹板测量的值。通过预制指夹板测量的参与者的近侧指间关节角度平均值为 16.55±7.90,通过定制夹板测量的所有参与者的近侧指间关节角度值均为“0”。通过预制指夹板和定制夹板测量的参与者的近侧指间关节角度值之间存在差异(p<0.001)。通过定制夹板测量的远侧指间关节角度值明显小于通过预制指夹板测量的值。
根据我们的研究,与预制指夹板相比,定制夹板可显著降低手指指间关节的弯曲度。