State Key Laboratory of Bioelectronics, School of Instrument Science and Engineering, Southeast University, Nanjing 210096, People's Republic of China.
Research Centre of Intelligent Healthcare, Coventry University, United Kingdom.
Physiol Meas. 2022 Sep 5;43(9). doi: 10.1088/1361-6579/ac890e.
: Accurate and reliable blood pressure (BP) measurement is important for the prevention and treatment of hypertension. The oscillometric-based automatic office blood pressure measurement (AOBPM) is widely used in hospitals and clinics, but measurement errors are common in BP measurements. There is a lack of systematic review of the sources of measurement errors.: A systematic review of all existing research on sources of AOBPM errors. A search strategy was designed in six online databases, and all the literature published before October 2021 was selected. Those studies that used the AOBPM device to measure BP from the upper arm of subjects were included.: A total of 1365 studies were screened, and 224 studies were included in this final review. They investigated 22 common error sources with clinical AOBPM. Regarding the causes of BP errors, this review divided them into the following categories: the activities before measurement, patient's factors, measurement environment, measurement procedure, and device settings. 13 sources caused increased systolic and diastolic BP (SBP and DBP), 2 sources caused the decrease in SBP and DBP, only 1 source had no significant effect on BPs, and the other errors had a non-uniform effect (either increase or decrease in BPs). The error ranges for SBP and DBP were -14 to 33 mmHg and -6 to 19 mmHg, respectively.: The measurement accuracy of AOBPM is susceptible to the influence of measurement factors. Interpreting BP readings need to be treated with caution in clinical measurements. This review made comprehensive evidence for the need for standardized BP measurements and provided guidance for clinical practitioners when measuring BP with AOBPM devices.
准确可靠的血压(BP)测量对于高血压的预防和治疗至关重要。基于示波法的自动诊室血压测量(AOBPM)广泛应用于医院和诊所,但在 BP 测量中常出现测量误差。目前缺乏对 AOBPM 测量误差来源的系统评价。
对所有关于 AOBPM 测量误差来源的现有研究进行系统评价。设计了一个在六个在线数据库中的搜索策略,并选择了截至 2021 年 10 月之前发表的所有文献。那些使用 AOBPM 设备从受试者上臂测量 BP 的研究被纳入。
共筛选出 1365 项研究,最终有 224 项研究纳入本综述。它们调查了 22 个常见的 AOBPM 临床误差源。关于 BP 误差的原因,本综述将其分为以下几类:测量前的活动、患者因素、测量环境、测量程序和设备设置。13 个来源导致收缩压和舒张压(SBP 和 DBP)升高,2 个来源导致 SBP 和 DBP 降低,只有 1 个来源对 BPs 没有显著影响,其他误差的影响不统一(SBP 和 DBP 升高或降低)。SBP 和 DBP 的误差范围分别为-14 至 33mmHg 和-6 至 19mmHg。
AOBPM 的测量准确性易受测量因素的影响。在临床测量中,解读 BP 读数需要谨慎。本综述为标准化 BP 测量提供了全面的证据,并为临床医生使用 AOBPM 设备测量 BP 时提供了指导。