Suppr超能文献

评估一种新的皮质条电极在大脑岛叶周围手术中术中体感监测的应用。

Evaluation of a new cortical strip electrode for intraoperative somatosensory monitoring during perirolandic brain surgery.

机构信息

Department of Neurosurgery, University Hospital, University of Zurich, Switzerland.

Department of Neurosurgery, Inselspital, Bern University Hospital, Bern, Switzerland.

出版信息

Clin Neurophysiol. 2022 Oct;142:44-51. doi: 10.1016/j.clinph.2022.07.497. Epub 2022 Aug 3.

Abstract

OBJECTIVE

During neurosurgical procedures, strip electrodes should have low impedance and sufficient adherence on the brain surface. We evaluated the signal quality, safety, and performance of a novel strip electrode (WISE Cortical Strip, WCS®), with conductive electrode contacts created with platinum nanoparticles embedded in a polymer base.

METHODS

In a multicenter interventional, non-inferiority study, we compared WCS to a conventional strip electrode (Ad-Tech). We recorded impedance and somatosensory evoked potentials (SEP) and determined the signal-to-noise ratio (SNR). We performed direct stimulation of the motor cortex. An external clinical event committee rated safety and adverse events and users rated usability.

RESULTS

During 32 brain surgeries in the paracentral region, WCS was rated safe and effective in signal transmission. Two seizure events were classified as probably related to the stimulation with WCS. The users rated WCS adhesion to the brain as satisfactory but reported difficulties sliding the WCS under the dura. The median (IQR) impedance of WCS was lower than for Ad-Tech: 2.7 (2.3-3.7) vs 5.30 (4.3-6.6) kΩ (p < 0.005). The SNR of SEP was non-inferior for WCS compared to Ad-Tech.

CONCLUSIONS

The impedance of WCS was lower than Ad-Tech without safety limitations. In small craniotomies not exposing the motor cortex its use may be limited.

SIGNIFICANCE

Low impedance electrodes facilitate recordings with high SNR.

摘要

目的

在神经外科手术中,条带电极应具有低阻抗和足够的脑表面附着力。我们评估了一种新型条带电极(WISE 皮质条带,WCS®)的信号质量、安全性和性能,其导电电极触点由嵌入聚合物基底中的铂纳米颗粒制成。

方法

在一项多中心介入、非劣效性研究中,我们将 WCS 与传统条带电极(Ad-Tech)进行了比较。我们记录了阻抗和体感诱发电位(SEP),并确定了信噪比(SNR)。我们对运动皮层进行直接刺激。一个外部临床事件委员会对安全性和不良事件进行了评级,用户对可用性进行了评级。

结果

在 32 例中央旁区脑手术中,WCS 被评为安全有效,信号传输可靠。有 2 例癫痫发作事件被归类为可能与 WCS 刺激有关。用户对 WCS 与大脑的附着力评价为满意,但报告在硬膜下滑动 WCS 时存在困难。WCS 的中位数(IQR)阻抗低于 Ad-Tech:2.7(2.3-3.7)比 5.30(4.3-6.6)kΩ(p<0.005)。与 Ad-Tech 相比,WCS 的 SEP SNR 不劣效。

结论

WCS 的阻抗低于 Ad-Tech,且没有安全限制。在不暴露运动皮层的小骨窗开颅术中,其使用可能受到限制。

意义

低阻抗电极有利于记录具有高 SNR 的信号。