Department of Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.
Oper Neurosurg (Hagerstown). 2022 Sep 1;23(3):254-260. doi: 10.1227/ons.0000000000000298. Epub 2022 Jun 14.
A number of stereotactic platforms are available for performing deep brain stimulation (DBS) lead implantation. Robot-assisted stereotaxy has emerged more recently demonstrating comparable accuracy and shorter operating room times compared with conventional frame-based systems.
To compare the accuracy of our streamlined robotic DBS workflow with data in the literature from frame-based and frameless systems.
We retrospectively reviewed 126 consecutive DBS lead placement procedures using a robotic stereotactic platform. Indications included Parkinson disease (n = 94), essential tremor (n = 21), obsessive compulsive disorder (n = 7), and dystonia (n = 4). Procedures were performed using a stereotactic frame for fixation and the frame pins as skull fiducials for robot registration. We used intraoperative fluoroscopic computed tomography for registration and postplacement verification.
The mean radial error for the target point was 1.06 mm (SD: 0.55 mm, range 0.04-2.80 mm) on intraoperative fluoroscopic computed tomography. The mean operative time for an asleep, bilateral implant without implantable pulse generator placement was 238 minutes (SD: 52 minutes), and skin-to-skin procedure time was 116 minutes (SD: 42 minutes).
We describe a streamlined workflow for DBS lead placement using robot-assisted stereotaxy with a comparable accuracy profile. Obviating the need for checking and switching coordinates, as is standard for frame-based DBS, also reduces the chance for human error and facilitates training.
有多种立体定向平台可用于进行深部脑刺激 (DBS) 导联植入。机器人辅助立体定向技术最近出现,与传统的基于框架的系统相比,其准确性相当,且手术室时间更短。
将我们简化的机器人 DBS 工作流程的准确性与基于框架和无框架系统的文献数据进行比较。
我们回顾性分析了使用机器人立体定向平台进行的 126 例连续 DBS 导联放置手术。适应证包括帕金森病 (n = 94)、特发性震颤 (n = 21)、强迫症 (n = 7)和肌张力障碍 (n = 4)。手术使用立体定向框架固定,框架销作为机器人注册的颅骨基准点。我们使用术中荧光透视计算机断层扫描进行注册和后置验证。
术中荧光透视计算机断层扫描时目标点的平均径向误差为 1.06 毫米 (SD:0.55 毫米,范围 0.04-2.80 毫米)。无植入式脉冲发生器放置的双侧植入物的平均手术时间为 238 分钟 (SD:52 分钟),皮肤到皮肤的手术时间为 116 分钟 (SD:42 分钟)。
我们描述了一种使用机器人辅助立体定向技术进行 DBS 导联放置的简化工作流程,其准确性相当。避免了像基于框架的 DBS 那样需要检查和切换坐标,从而减少了人为错误的机会,并便于培训。