Department of Surgery, Meander Medical Centre, Maatweg 3, 3813 TZ, Amersfoort, The Netherlands.
Technical Medicine, University of Twente, Enschede, The Netherlands.
Eur J Orthop Surg Traumatol. 2023 Jul;33(5):1797-1804. doi: 10.1007/s00590-022-03350-8. Epub 2022 Aug 17.
The Gamma3 nail (Stryker®) is an intramedullary device consisting of a proximal lag screw and distal interlocking screw. It is still unknown whether the screw locking mode could influence clinical outcomes. The aim of this study is to compare static and dynamic screw locking regarding their influence on surgical revisions and lag screw displacement.
A retrospective single-centre study was performed on patients ≥ 60 years admitted for a trochanteric fracture between September 2016 and January 2020. Surgical revisions and lag screw displacement were evaluated at 6 weeks and 1-year follow-up, respectively.
A total of 142 patients were included for analysis. Surgical revisions were needed in 13 cases (9.2%). Indications included implant breakage (n = 3), lag screw cut-out (n = 3), lateral hip pain (n = 6) and non-union (n = 1). The number of surgical revisions was not different between static and dynamic locking (OR 2.55; 95%CI 0.73-8.56; p 0.142). The median lag screw displacement was 2.5 mm, which was similar for static and dynamic locking (2.3 mm versus 2.7 mm; p 0.785).
The screw locking mode of the Gamma3 nail is not associated with a higher risk of surgical revisions. However, the design of the Gamma3 nail may not be suitable for static locking.
Gamma3 钉(Stryker®)是一种由近端锁定螺钉和远端锁定螺钉组成的髓内装置。目前尚不清楚螺钉锁定方式是否会影响临床结果。本研究旨在比较静态和动态锁定对手术翻修和锁定螺钉移位的影响。
对 2016 年 9 月至 2020 年 1 月期间因转子间骨折入院的年龄≥60 岁的患者进行了回顾性单中心研究。分别在术后 6 周和 1 年随访时评估手术翻修和锁定螺钉移位情况。
共纳入 142 例患者进行分析。13 例(9.2%)需要手术翻修。翻修指征包括植入物断裂(n=3)、锁定螺钉切出(n=3)、髋关节外侧疼痛(n=6)和骨不连(n=1)。静态和动态锁定的手术翻修次数无差异(OR 2.55;95%CI 0.73-8.56;p=0.142)。锁定螺钉的中位数移位为 2.5mm,静态和动态锁定相似(2.3mm 与 2.7mm;p=0.785)。
Gamma3 钉的螺钉锁定方式与更高的手术翻修风险无关。然而,Gamma3 钉的设计可能不适合静态锁定。