Yoon Sumi, Hur Mina, Kim Hanah, Moon Hee-Won, Yun Yeo-Min
Department of Laboratory Medicine, Chung-Ang University College of Medicine, Seoul 06973, Korea.
Department of Laboratory Medicine, Konkuk University School of Medicine, Seoul 05030, Korea.
Diagnostics (Basel). 2022 Aug 10;12(8):1930. doi: 10.3390/diagnostics12081930.
The newly developed Axis-Shield clinical chemistry heparin-binding protein (HBP) assay (Axis-Shield Diagnostics Ltd., Dundee, Scotland) can be applied to fully automated platforms. We aimed to establish a reference interval (RI) of HBP using the Axis-Shield HBP assay, and to evaluate the analytical performance of this assay. An RI was established in 212 sodium citrated plasma samples using the non-parametric method (2.5th and 97.5th percentiles). Precision, linearity, and carry-over were evaluated according to the Clinical and Laboratory Standards Institute guidelines. The RI of HBP was between 5.3 ng/mL and 171.0 ng/mL, which could be applied regardless of gender and age. Percentage coefficients of variations (%CVs) of repeatability and within-laboratory precision were 4.9% and 6.3%, respectively, for low-concentration control and 1.6% and 3.0%, respectively, for high-concentration control. The linearity was excellent (coefficient of determination (R) = 0.99), and the carry-over rate was negligible (0.05%). This is the first study to establish an RI of HBP using the newly developed and fully automated Axis-Shield HBP assay. The Axis-Shield HBP assay showed an acceptable level of analytical performance and could be used to measure HBP concentrations effectively in routine clinical practice. Further studies are awaited to evaluate the clinical utility of HBP using this automated assay.
新开发的Axis-Shield临床化学肝素结合蛋白(HBP)检测法(Axis-Shield诊断有限公司,苏格兰邓迪)可应用于全自动平台。我们旨在使用Axis-Shield HBP检测法建立HBP的参考区间(RI),并评估该检测法的分析性能。使用非参数方法(第2.5百分位数和第97.5百分位数)在212份枸橼酸钠血浆样本中建立了RI。根据临床和实验室标准协会指南评估精密度、线性和携带污染。HBP的RI在5.3 ng/mL至171.0 ng/mL之间,无论性别和年龄均可适用。低浓度质控品的重复性和实验室内精密度的变异系数百分比(%CVs)分别为4.9%和6.3%,高浓度质控品分别为1.6%和3.0%。线性良好(决定系数(R)=0.99),携带污染率可忽略不计(0.05%)。这是第一项使用新开发的全自动Axis-Shield HBP检测法建立HBP的RI的研究。Axis-Shield HBP检测法显示出可接受的分析性能水平,可用于在常规临床实践中有效测量HBP浓度。有待进一步研究以评估使用这种自动化检测法检测HBP的临床效用。