The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M.
A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3.
There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA.
It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other.
生物标志物的值在常规临床决策中起着核心作用。然而,不同自动化化学分析仪的性能仍不明确。为了确定不同平台的性能,我们比较了罗氏 Cobas 8000 和迈瑞 BS2000M 之间的一致性和准确性。
共收集了 1869 份剩余血清样本。根据 CLSI EP5-A3,使用配对 t 检验、Passing-Bablok 回归分析和 Bland-Altman 分析对肌酸激酶(CK)、乳酸脱氢酶-1(LDH-1)、视黄醇结合蛋白(RBP)、胱抑素 C(Cys-C)、免疫球蛋白 A(IgA)、免疫球蛋白 M(IgM)和免疫球蛋白 G(IgG)进行评估。
两台机器之间所有项目的平均偏差存在显著差异(P<0.001)。由于截距 A 的 95%置信区间包含 0,CK、LDH-1、Cys-C 和 IgG 在 Passing-Bablok 回归分析中未显示系统误差。IgM 中斜率 B 的 95%置信区间包含 1,IgM 的两次测量无差异。除 IgA 外,所有项目的 r 值和相关系数均高于 0.91,表明相关性和一致性良好。Bland-Altman 分析表明,除 CK、LDH-1 和 IgA 外,两台仪器的点有超过 95%的差异。
可以认为两台仪器在 CK、LDH-1、RBP、Cys-C、IgM 和 IgG 方面具有良好的相关性和一致性,两台仪器可互换且能相互替代。