两种自动生化分析仪系统的性能评估:罗氏Cobas 8000和迈瑞BS2000M。

Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M.

作者信息

Chen Mingxing, Qin Simeng, Yang Sitao, Chen Huaping, Lu Liuyi, Qin Xue

机构信息

First Affiliated Hospital of Guangxi Medical University, Department of Clinical Laboratory, China.

出版信息

J Med Biochem. 2022 Jul 29;41(3):306-315. doi: 10.5937/jomb0-34328.

Abstract

BACKGROUND

The values of biomarkers play a central role in routine clinical decision-making. Whereas the performance of different automated chemical analyzers remains unclear. To determine the performance of different platforms, we compared the consistency and accuracy between Roche Cobas 8000 and Mindray BS2000M.

METHODS

A total of 1869 remaining serum samples were collected. CK, LDH-1, RBP, Cys-C, IgA, IgM, and IgG were assessed using paired t-test, Passing-Bablok regression analysis, and Bland-Altman analysis according to CLSI EP5-A3.

RESULTS

There were significant differences in the average bias of all items between the two machines (P<0.001). Because the 95% confidence interval of intercept A included 0, CK, LDH-1, Cys-C and IgG did not show systematic error in Passing-Bablok regression analysis. The confidence interval of 95% of the slope B in IgM contained 1, and there was no difference in the two measurements in IgM. Except for IgA, the r values and correlation coefficient of all items were higher than 0.91, which showed that the correlation and consistency were good. The Bland-Altman analysis showed that two instruments had more than 95% of the points apart from CK, LDH-1, and IgA.

CONCLUSIONS

It can be considered that the two instruments have good correlation and consistency in CK, LDH-1, RBP, Cys-C, IgM, and IgG, and the two instruments are interchangeable and can replace each other.

摘要

背景

生物标志物的值在常规临床决策中起着核心作用。然而,不同自动化化学分析仪的性能仍不明确。为了确定不同平台的性能,我们比较了罗氏 Cobas 8000 和迈瑞 BS2000M 之间的一致性和准确性。

方法

共收集了 1869 份剩余血清样本。根据 CLSI EP5-A3,使用配对 t 检验、Passing-Bablok 回归分析和 Bland-Altman 分析对肌酸激酶(CK)、乳酸脱氢酶-1(LDH-1)、视黄醇结合蛋白(RBP)、胱抑素 C(Cys-C)、免疫球蛋白 A(IgA)、免疫球蛋白 M(IgM)和免疫球蛋白 G(IgG)进行评估。

结果

两台机器之间所有项目的平均偏差存在显著差异(P<0.001)。由于截距 A 的 95%置信区间包含 0,CK、LDH-1、Cys-C 和 IgG 在 Passing-Bablok 回归分析中未显示系统误差。IgM 中斜率 B 的 95%置信区间包含 1,IgM 的两次测量无差异。除 IgA 外,所有项目的 r 值和相关系数均高于 0.91,表明相关性和一致性良好。Bland-Altman 分析表明,除 CK、LDH-1 和 IgA 外,两台仪器的点有超过 95%的差异。

结论

可以认为两台仪器在 CK、LDH-1、RBP、Cys-C、IgM 和 IgG 方面具有良好的相关性和一致性,两台仪器可互换且能相互替代。