科华Polaris c2000全自动生化分析仪自动分析仪室内质量控制的功能评估

Function evaluation of the autoanalyser internal quality control of the Kehua Polaris c2000 automatic biochemical analyser.

作者信息

Wang Yiyi, Shi Lei, Zhang Jue, Chen Chen

机构信息

Shanghai Shuguang Hospital Affiliated with Shanghai University of TCM, Department of Laboratory Medicine, Shanghai, China.

出版信息

J Med Biochem. 2024 Nov 16;43(6):853-859. doi: 10.5937/jomb0-47151.

Abstract

BACKGROUND

The present study aims to (1) evaluate the autoanalyser quality control (QC) module of the Kehua Polaris c2000 automatic modularised biochemical analyser and (2) verify the impact of the analyser's automatic implementation of internal QC (IQC) testing at a set time point on the results of IQC and turnaround time (TAT).

METHODS

For 5 consecutive days, three different methods were used to conduct IQC. Method 1: Internal QC was carried out at 8:00 every day. Method 2: The QC products were placed in the instrument calibration QC plate in the afternoon, and the instrument was set to measure the products at 7:00 the next day automatically. Method 3: The QC products were placed in the instrument calibration QC plate, and the instrument was set up for automatic measurement at 7:30 every day. All three methods were compared and evaluated. The effect of IQC on the TAT was monitored using method 2.

RESULTS

There were no statistical differences in the IQC results between methods 2 and 1. However, there were statistical differences in some items between methods 3 and 1; thus, the IQC results of method 2 can be adopted. Implementing method 2 for IQC can help achieve a significant TAT-saving of 35.23 min during the automatic retest process after the IQC indicates an out-of-control situation. This time reduction is highly valuable for improving efficiency and streamlining the testing workflow.

CONCLUSIONS

Using the autoanalyser QC module of the Kehua Polaris c2000 automatic modular biochemical analysis system to perform IQC has no impact on the IQC test results and can save TAT, as well as automatically correct most out-of-control occurrences.

摘要

背景

本研究旨在(1)评估科华北极星c2000全自动模块化生化分析仪的自动分析仪质量控制(QC)模块,以及(2)验证分析仪在设定时间点自动实施内部质量控制(IQC)测试对IQC结果和周转时间(TAT)的影响。

方法

连续5天,采用三种不同方法进行IQC。方法1:每天8:00进行内部QC。方法2:下午将QC产品放置在仪器校准QC板中,仪器设置为第二天7:00自动测量这些产品。方法3:将QC产品放置在仪器校准QC板中,仪器设置为每天7:30自动测量。对这三种方法进行比较和评估。使用方法2监测IQC对TAT的影响。

结果

方法2和方法1的IQC结果无统计学差异。然而,方法3和方法1在某些项目上存在统计学差异;因此,可以采用方法2的IQC结果。在IQC显示失控情况后的自动重新测试过程中,采用方法2进行IQC可显著节省35.23分钟的TAT。这一时间减少对于提高效率和简化测试工作流程非常有价值。

结论

使用科华北极星c2000自动模块化生化分析系统的自动分析仪QC模块进行IQC对IQC测试结果没有影响,并且可以节省TAT,还能自动纠正大多数失控情况。