The purpose of this study was to determine the effect of rotation and tilt on the radiographic teardrop distance (TD) on anteroposterior (AP) pelvis radiographs.
Radiographic examination of a pelvis models was conducted utilizing increasing degrees of beam rotation and tilt on portable C-arm fluoroscopy. The TD, x-plane rotation (symphyseal-mid-sacrum distance (SMS)), and y-plane tilt (sacroiliac joint-symphysis distance (SIS)) were measured by four independent observers. Interobserver reliability was assessed using intraclass correlations.
TD was altered by less than 2 mm with up to 7.5° fluoroscopic rotation (SMS: 3 cm) and up to 30° of inlet and 15° of outlet (SIS: ± 3.3 cm). SMS distance effectively corresponded to the degree of rotation present (r = 1.00, CI: 0.97 to 1.00, p < 0.0001) and was strongly correlated to TD (r = -0.95, CI: -0.99 to -0.67, p = 0.001). SIS distance effectively corresponded to the degree of tilt present (r = -0.97, CI: -0.99 to -0.88, p < 0.0001) and was correlated to TD (r = 0.94, CI: 0.75 to 0.99, p = 0.0001). Linear regression models determined that, with every degree of rotation and tilt, TD was altered by 0.4 mm and 0.09 mm, respectively (p = 0.0004, r = 0.93 and p < 0.0001, r = 0.94, respectively). Interobserver reliability among observers was excellent (0.92).
The TD has excellent interobserver reliability and is minimally impacted by up to 7.5° of rotation, 30° inlet tilt, and 15° of outlet tilt. Utilization of these thresholds may ensure reliability of TD measurements when assessing pelvis stress radiographs.
本研究旨在确定旋转和倾斜对前后位(AP)骨盆 X 线片上泪滴距离(TD)的影响。
在便携式 C 臂透视机上进行骨盆模型的 X 线检查,增加射线的旋转和倾斜角度。由四位独立观察者测量 TD、X 平面旋转(耻骨联合-中骶骨距离(SMS))和 Y 平面倾斜(骶髂关节-耻骨联合距离(SIS))。采用组内相关系数评估观察者间的可靠性。
TD 的变化小于 2mm,旋转角度不超过 7.5°(SMS:3cm),入口倾斜不超过 30°,出口倾斜不超过 15°(SIS:±3.3cm)。SMS 距离与存在的旋转角度有效对应(r=1.00,CI:0.97 至 1.00,p<0.0001),与 TD 高度相关(r=-0.95,CI:-0.99 至 -0.67,p=0.001)。SIS 距离与存在的倾斜角度有效对应(r=-0.97,CI:-0.99 至 -0.88,p<0.0001),与 TD 相关(r=0.94,CI:0.75 至 0.99,p=0.0001)。线性回归模型确定,每增加 1 度旋转和倾斜,TD 分别改变 0.4mm 和 0.09mm(p=0.0004,r=0.93 和 p<0.0001,r=0.94)。观察者之间的观察者间可靠性极好(0.92)。