The effect of pelvic ring rotation and tilt on the radiographic teardrop distance: an important consideration in the assessment of dynamic displacement on stress radiographs.

机构信息

Department of Orthopedics, Denver Health Medical Center, 777 Bannock St, MC 0188, Denver, CO, 80204, USA.

Department of Orthopedics, University of Colorado School of Medicine, Aurora, CO, USA.

出版信息

Eur J Orthop Surg Traumatol. 2023 Jul;33(5):1953-1957. doi: 10.1007/s00590-022-03377-x. Epub 2022 Sep 1.

Abstract

PURPOSE

The purpose of this study was to determine the effect of rotation and tilt on the radiographic teardrop distance (TD) on anteroposterior (AP) pelvis radiographs.

METHODS

Radiographic examination of a pelvis models was conducted utilizing increasing degrees of beam rotation and tilt on portable C-arm fluoroscopy. The TD, x-plane rotation (symphyseal-mid-sacrum distance (SMS)), and y-plane tilt (sacroiliac joint-symphysis distance (SIS)) were measured by four independent observers. Interobserver reliability was assessed using intraclass correlations.

RESULTS

TD was altered by less than 2 mm with up to 7.5° fluoroscopic rotation (SMS: 3 cm) and up to 30° of inlet and 15° of outlet (SIS: ± 3.3 cm). SMS distance effectively corresponded to the degree of rotation present (r = 1.00, CI: 0.97 to 1.00, p < 0.0001) and was strongly correlated to TD (r = -0.95, CI: -0.99 to -0.67, p = 0.001). SIS distance effectively corresponded to the degree of tilt present (r = -0.97, CI: -0.99 to -0.88, p < 0.0001) and was correlated to TD (r = 0.94, CI: 0.75 to 0.99, p = 0.0001). Linear regression models determined that, with every degree of rotation and tilt, TD was altered by 0.4 mm and 0.09 mm, respectively (p = 0.0004, r = 0.93 and p < 0.0001, r = 0.94, respectively). Interobserver reliability among observers was excellent (0.92).

CONCLUSION

The TD has excellent interobserver reliability and is minimally impacted by up to 7.5° of rotation, 30° inlet tilt, and 15° of outlet tilt. Utilization of these thresholds may ensure reliability of TD measurements when assessing pelvis stress radiographs.

摘要

目的

本研究旨在确定旋转和倾斜对前后位(AP)骨盆 X 线片上泪滴距离(TD)的影响。

方法

在便携式 C 臂透视机上进行骨盆模型的 X 线检查,增加射线的旋转和倾斜角度。由四位独立观察者测量 TD、X 平面旋转(耻骨联合-中骶骨距离(SMS))和 Y 平面倾斜(骶髂关节-耻骨联合距离(SIS))。采用组内相关系数评估观察者间的可靠性。

结果

TD 的变化小于 2mm,旋转角度不超过 7.5°(SMS:3cm),入口倾斜不超过 30°,出口倾斜不超过 15°(SIS:±3.3cm)。SMS 距离与存在的旋转角度有效对应(r=1.00,CI:0.97 至 1.00,p<0.0001),与 TD 高度相关(r=-0.95,CI:-0.99 至 -0.67,p=0.001)。SIS 距离与存在的倾斜角度有效对应(r=-0.97,CI:-0.99 至 -0.88,p<0.0001),与 TD 相关(r=0.94,CI:0.75 至 0.99,p=0.0001)。线性回归模型确定,每增加 1 度旋转和倾斜,TD 分别改变 0.4mm 和 0.09mm(p=0.0004,r=0.93 和 p<0.0001,r=0.94)。观察者之间的观察者间可靠性极好(0.92)。

结论

TD 具有极好的观察者间可靠性,受最大 7.5°的旋转、30°入口倾斜和 15°出口倾斜的影响极小。在评估骨盆应力量线片时,利用这些阈值可确保 TD 测量的可靠性。