Suppr超能文献

AliveCor KardiaMobile六导联心电图与标准心电图在儿科患者中的比较

Comparison of AliveCor KardiaMobile Six-Lead ECG with Standard ECG in Pediatric Patients.

作者信息

Girvin Zachary P, Silver Eric S, Liberman Leonardo

机构信息

Vagelos College of Physicians and Surgeons, Columbia University, New York, NY, USA.

Department of Pediatrics, Columbia University Irving Medical Center, 3959 Broadway Ave - 2 North, New York, NY, 10032, USA.

出版信息

Pediatr Cardiol. 2023 Mar;44(3):689-694. doi: 10.1007/s00246-022-02998-7. Epub 2022 Sep 3.

Abstract

The AliveCor KardiaMobile (ACKM) is a remote electrocardiogram (ECG) monitoring device. Little research has been conducted on its accuracy with pediatric patients. This prospective study aims to compare the ACKM six-lead device with a standard fifteen-lead ECG in measuring the QTc, QRS, and axis in pediatric patients. Pediatric patients ages 5 to 21 years were enrolled prospectively to have their ECG recorded using an ACKM six-lead device following a recording with the standard 15-lead ECG. A pediatric electrophysiologist measured the QTc, QRS interval, and QRS axis for both ECGs. Bland-Altman analysis was performed to assess agreement among measurements. The study included 141 patients. The mean age was 12.3 ± 4.4 years. Average heart rate was 79 ± 16 bpm. The mean difference in the QTc measurements for a paired standard ECG and ACKM was - 0.6 ms [95% confidence interval - 48 to 47 ms]. Of the ACKM QTc measurements, 117 (83%) were within 30 ms of the standard ECG. The mean difference in paired QRS measurements was - 1.3 ms [95% confidence interval - 23 to 21 ms]. Of the ACKM QRS measurements, 134 (95%) were within 20 ms of the standard ECG. The measured axis was the same for 84% of ACKM and standard ECGs. Over 80% of the ACKM six-lead ECGs produced QTc, QRS, and axis deviation measurements within a clinically useful range of the standard ECG. However, it is not accurate enough to be used consistently in place of a standard ECG for QTc and QRS measurement for pediatric patients.

摘要

AliveCor KardiaMobile(ACKM)是一种远程心电图(ECG)监测设备。针对其在儿科患者中的准确性,相关研究较少。这项前瞻性研究旨在比较ACKM六导联设备与标准十五导联心电图在测量儿科患者QTc、QRS和电轴方面的差异。前瞻性纳入了年龄在5至21岁的儿科患者,先使用标准15导联心电图进行记录,随后使用ACKM六导联设备记录其心电图。由一名儿科电生理学家测量两种心电图的QTc、QRS间期和QRS电轴。采用Bland-Altman分析评估测量结果之间的一致性。该研究共纳入141例患者。平均年龄为12.3±4.4岁。平均心率为79±16次/分钟。配对的标准心电图和ACKM在QTc测量上的平均差异为-0.6毫秒[95%置信区间-48至47毫秒]。在ACKM的QTc测量中,117例(83%)在标准心电图的30毫秒范围内。配对QRS测量的平均差异为-1.3毫秒[95%置信区间-23至21毫秒]。在ACKM的QRS测量中,134例(95%)在标准心电图的20毫秒范围内。84%的ACKM和标准心电图测量的电轴相同。超过80%的ACKM六导联心电图在标准心电图临床有用范围内产生QTc、QRS和电轴偏差测量值。然而,对于儿科患者的QTc和QRS测量,其准确性不足以始终替代标准心电图使用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验