不同口腔抽吸装置在超声洁牙和高速手机使用过程中的气溶胶减少效果。

Aerosol reduction efficacy of different intra-oral suction devices during ultrasonic scaling and high-speed handpiece use.

机构信息

Oral Sciences, Glasgow Dental Hospital and School, College of Medical, Veterinary and Life Sciences, University of Glasgow, Glasgow, G2 3JZ, UK.

Global Clinical Affairs, Dentsply Sirona, York, PA, USA.

出版信息

BMC Oral Health. 2022 Sep 6;22(1):388. doi: 10.1186/s12903-022-02386-w.

Abstract

BACKGROUND

The COVID-19 pandemic led to significant changes in the provision of dental services, aimed at reducing the spread of respiratory pathogens through restrictions on aerosol generating procedures (AGPs). Evaluating the risk that AGPs pose in terms of SARS-CoV-2 transmission is complex, and measuring dental aerosols is challenging. To date, few studies focus on intra-oral suction. This study sought to assess the effectiveness of commonly used intra-oral suction devices on aerosol mitigation.

METHODS

Ultrasonic scaling and high-speed handpiece procedures were undertaken to generate aerosol particles. Multiple particle sensors were positioned near the oral cavity. Sensor data were extracted using single board computers with custom in-house Bash code. Different high-volume and low-volume suction devices, both static and dynamic, were evaluated for their efficacy in preventing particle escape during procedures.

RESULTS

In all AGPs the use of any suction device tested resulted in a significant reduction in particle counts compared with no suction. Low-volume and static suction devices showed spikes in particle count demonstrating moments where particles were able to escape from the oral cavity. High-volume dynamic suction devices, however, consistently reduced the particle count to background levels, appearing to eliminate particle escape.

CONCLUSIONS

Dynamic high-volume suction devices that follow the path of the aerosol generating device effectively eliminate aerosol particles escaping from the oral cavity, in contrast to static devices which allow periodic escape of aerosol particles. Measuring the risk of SARS-CoV-2 transmission in a dental setting is multi-factorial; however, these data suggest that the appropriate choice of suction equipment may further reduce the risk from AGPs.

摘要

背景

COVID-19 大流行导致牙科服务发生重大变化,旨在通过限制产生气溶胶的程序(AGP)来减少呼吸道病原体的传播。评估 AGP 对 SARS-CoV-2 传播的风险是复杂的,测量牙科气溶胶具有挑战性。迄今为止,很少有研究关注口腔内抽吸。本研究旨在评估常用口腔内抽吸设备对气溶胶缓解的效果。

方法

使用超声洁牙和高速手机程序产生气溶胶颗粒。多个粒子传感器放置在口腔附近。使用带有自定义 Bash 代码的单板计算机提取传感器数据。评估了多种高容量和低容量抽吸设备(静态和动态)在程序过程中防止颗粒逸出的效果。

结果

在所有 AGP 中,与无抽吸相比,使用任何抽吸设备都会导致颗粒计数显著减少。低容量和静态抽吸设备的颗粒计数出现峰值,表明颗粒有从口腔逸出的时刻。然而,高容量动态抽吸设备始终将颗粒计数降低到背景水平,似乎消除了颗粒逸出。

结论

与允许气溶胶颗粒周期性逸出的静态设备相比,遵循气溶胶生成设备路径的动态高容量抽吸设备可有效消除从口腔逸出的气溶胶颗粒。在牙科环境中测量 SARS-CoV-2 传播的风险是多方面的;然而,这些数据表明,适当选择抽吸设备可能会进一步降低 AGP 的风险。