Radiography (Lond). 2022 Nov;28(4):1116-1121. doi: 10.1016/j.radi.2022.08.006. Epub 2022 Sep 10.
The detectability of low contrast detail (LCD) is a method used to assess image quality (IQ) in neonatal radiography; however, there is a lack of data on the relationship between LCD detectability and visual IQ. The study aims at investigating the relationship between the LCD detectability and visual IQ and pathology visibility (PV).
Several acquisition parameters were employed to obtain a group of images from a neonatal Gammex chest phantom. Three observers applied relative visual grading analysis (VGA) for assessing the IQ and PV. A simulated pneumothorax visibility (PNV) and simulated hyaline membrane disease visibility (HMV) represented PV. Next, a CDRAD 2.0 phantom was radiographed utilising the same acquisition protocols, and several paired images were obtained. With the use of CDRAD analyser software, the detectability of LCD was assessed and expressed by an image quality figure inverse (IQFi) metric. The correlation between the IQF and each of IQ, PNV and HMV was examined.
The physical measure (IQF) and the visual assessment of IQ were shown to be strongly correlated (r = 0.95; p < 0.001). Using Pearson's correlation, the IQF, PNV, and HMV were found to be strongly correlated (r = 0.94; p < 0.001) and (r = 0.92; p < 0.001), correspondingly.
Results of the study show that physical measures of LCD detectability utilising the CDRAD 2.0 phantom is strongly corelated with visual IQ and PV (PNV and HMV) and can be used to evaluate IQ when undertaking neonatal chest radiography (CXR).
This study establishes the feasibility of utilising the physical measure (IQF) and the CDRAD 2.0 phantom in routine quality assurance and neonatal CXR optimisation studies.
低对比度细节(LCD)的检测能力是评估新生儿放射摄影图像质量(IQ)的一种方法;然而,关于 LCD 检测能力与视觉智商之间的关系的数据却很少。本研究旨在探讨 LCD 检测能力与视觉智商和病理学可见性(PV)之间的关系。
使用几种采集参数从新生儿 Gammex 胸部体模中获取一组图像。三名观察者应用相对视觉分级分析(VGA)评估 IQ 和 PV。模拟气胸可见性(PNV)和模拟透明膜病可见性(HMV)代表 PV。接下来,使用相同的采集协议对 CDRAD 2.0 体模进行拍摄,并获得了几组配对图像。利用 CDRAD 分析软件评估 LCD 的检测能力,并以图像质量因子倒数(IQFi)的度量表示。检查 IQF 与 IQ、PNV 和 HMV 之间的相关性。
物理测量(IQF)和 IQ 的视觉评估显示出很强的相关性(r=0.95;p<0.001)。使用 Pearson 相关系数,发现 IQF、PNV 和 HMV 之间存在很强的相关性(r=0.94;p<0.001)和(r=0.92;p<0.001)。
研究结果表明,使用 CDRAD 2.0 体模进行的 LCD 检测能力的物理测量与视觉 IQ 和 PV(PNV 和 HMV)具有很强的相关性,可以在进行新生儿胸部 X 线摄影(CXR)时用于评估 IQ。
本研究确立了在常规质量保证和新生儿 CXR 优化研究中使用物理测量(IQF)和 CDRAD 2.0 体模的可行性。