Abstract

BACKGROUND

This study aimed to investigate improvements in Ga-FAPI PET/CT image quality due to using total variation regularized expectation maximization (TVR-EM) and ordered subset expectation maximization (OS-EM) reconstruction.

METHODS

Data from a total of 24 patients were retrospectively analyzed in this study. Positron emission tomography (PET) images were reconstructed using OS-EM and TVR-EM for 2 and 3 minutes-per-bed (min/bed) acquisition. The SUVmean of a region-of-interest on the liver, image noise, signal-to-noise ratios (SNR), the SUVmax of the lesions and the tumour-to-background ratios (TBR) were measured and compared between the 2 methods. Subjective image qualities were evaluated by two experienced radiologists using a 5-point score scale (5-excellent, 1-poor).

RESULTS

In total, 132 lesions were analyzed. The image noise in TVR-EM reconstruction groups was lower than in the OS-EM groups (all P<0.05). The SNR, SUVmax of lesions and the TBR were higher for the TVR-EM reconstruction groups compared to OS-EM groups (all P<0.05). Also, the SUVmax of the lesions in the TVR-EM groups increased by at least 12% compared to OS-EM 3 min/bed group. The SUVmax for small (<10 mm) and large lesions (>10 mm) in the TVR-EM 2 min/bed group were significantly larger compared to the OS-EM 3 min/bed groups (all P<0.05). The highest image quality score resulted from the TVR-EM 3 min/bed group with a penalization factor of 0.25 (3.92±0.19).

CONCLUSIONS

TVR-EM reduces image noise and improved the SNR, SUVmax and TBR of the lesions. It also enables fast acquisition without compromising image quality compared to standard OS-EM.

摘要

背景

本研究旨在探讨全变分正则化期望最大化(TVR-EM)和有序子集期望最大化(OS-EM)重建对 Ga-FAPI PET/CT 图像质量的改善。

方法

本研究回顾性分析了 24 例患者的数据。使用 OS-EM 和 TVR-EM 对 2 分钟/床位(min/bed)和 3 分钟/床位采集的 PET 图像进行重建。测量并比较了两种方法的肝脏感兴趣区(ROI)的 SUVmean、图像噪声、信噪比(SNR)、病灶 SUVmax和肿瘤与背景比值(TBR)。两位有经验的放射科医生使用 5 分制(5-优秀,1-差)对主观图像质量进行评估。

结果

共分析了 132 个病灶。TVR-EM 重建组的图像噪声低于 OS-EM 组(均 P<0.05)。与 OS-EM 组相比,TVR-EM 重建组的 SNR、病灶 SUVmax和 TBR 更高(均 P<0.05)。此外,与 OS-EM 3 min/bed 组相比,TVR-EM 组的病灶 SUVmax至少增加了 12%。TVR-EM 2 min/bed 组中小(<10mm)和大(>10mm)病灶的 SUVmax均显著大于 OS-EM 3 min/bed 组(均 P<0.05)。具有 0.25 惩罚因子的 TVR-EM 3 min/bed 组获得了最高的图像质量评分(3.92±0.19)。

结论

TVR-EM 降低了图像噪声,提高了病灶的 SNR、SUVmax和 TBR。与标准 OS-EM 相比,它还可以在不影响图像质量的情况下实现快速采集。