Porubszky T
J Biomed Eng. 1987 Jul;9(3):261-6. doi: 10.1016/0141-5425(87)90011-2.
Explicit expressions are given for the relative variance of the output screen of X-ray image intensifiers, and for detection quantum efficiency as a function of the technical features of the intensifier. The analysis takes into account all conversion stages and physical processes inside the image intensifier and includes a special treatment of scintillation. The results are analysed numerically both for previously reported approximations and for applying this more recent method of detailed visibility calculations.