Rowlands J A, Taylor K W
Med Phys. 1983 Nov-Dec;10(6):786-95. doi: 10.1118/1.595357.
The measured and theoretically predicted values of detective quantum efficiency (DQE) for a CsI x-ray image intensifier are compared for nine monoenergetic beams of x rays. The agreement between measurement and theory of better than +/- 5% indicates that we have a sound understanding of the physical parameters controlling the DQE. It is shown that the fraction of K-fluorescent x rays escaping from the input phosphor is independent of incident energy. The number of electrons released within the x-ray image intensifier (XRII) by an incident x ray has been measured. The mechanism for energy broadening within the XRII is shown to be predominantly the limited number of electrons and not light absorption.
针对九束单能X射线束,比较了碘化铯X射线影像增强器的探测量子效率(DQE)的测量值和理论预测值。测量值与理论值之间优于±5%的一致性表明,我们对控制DQE的物理参数有了充分的理解。结果表明,从输入荧光体逸出的K系荧光X射线的比例与入射能量无关。测量了入射X射线在X射线影像增强器(XRII)内释放的电子数量。结果表明,XRII内能量展宽的机制主要是电子数量有限,而非光吸收。