MAGMA. 2023 Feb;36(1):15-23. doi: 10.1007/s10334-022-01042-2. Epub 2022 Sep 27.
Recent work showed the feasibility of measuring velocity pulsatility in the perforating arteries at the level of the BG using 3T MRI. However, test-retest measurements have not been performed, yet. This study assessed the test-retest reliability of 3T MRI blood flow velocity measurements in perforating arteries in the BG.
Two-dimensional phase-contrast cardiac gated (2D-PC) images were acquired for 35 healthy controls and repeated with and without repositioning. 2D-PC images were processed and analyzed, to assess the number of detected perforating arteries (N), mean blood flow velocity (V), and velocity pulsatility index (vPI). Paired t-tests and Bland-Altman plots were used to compare variance in outcome parameters with and without repositioning, and limits of agreement (LoA) were calculated.
The LoA was smallest for V (35%) and highest for vPI (79%). Test-retest reliability was similar with and without repositioning of the subject.
We found similar LoA with and without repositioning indicating that the measurement uncertainty is dominated by scanner and physiological noise, rather than by planning. This enables to study hemodynamic parameters in perforating arteries at clinically available scanners, provided sufficiently large sample sizes are used to mitigate the contribution of scanner- and physiological noise.
最近的研究表明,使用 3T MRI 可以在 BG 水平测量穿支动脉的速度波动。然而,尚未进行测试-重测测量。本研究评估了 3T MRI 在 BG 穿支动脉中测量血流速度的测试-重测可靠性。
对 35 名健康对照者进行二维相位对比心脏门控(2D-PC)图像采集,并在重新定位和不重新定位的情况下重复采集。处理和分析 2D-PC 图像,以评估检测到的穿支动脉数量(N)、平均血流速度(V)和速度搏动指数(vPI)。使用配对 t 检验和 Bland-Altman 图比较重新定位前后各参数的方差,并计算差异限度(LoA)。
vPI 的 LoA 最高(79%),V 的 LoA 最小(35%)。无论是否重新定位受检者,测试-重测的可靠性均相似。
我们发现重新定位前后的 LoA 相似,表明测量的不确定性主要由扫描仪和生理噪声决定,而不是由规划决定。这使得可以在临床可用的扫描仪上研究穿支动脉的血流动力学参数,只要使用足够大的样本量来减轻扫描仪和生理噪声的影响。