Vet Rec. 2022 Dec;191(12):e2184. doi: 10.1002/vetr.2184. Epub 2022 Oct 5.
Electrical impedance tomography (EIT) produces lung ventilation images via a thoracic electrode belt. Robust electrode design and material, providing low electrode skin contact impedance (SCI), is needed in veterinary medicine. The aim of this study was to compare three EIT electrode designs and materials.
Simulations of cylindrical, rectangular and spiked electrode designs were used to evaluate electrode SCI as a function of electrode size, where skin contact was uneven. Gold-plated washers (E ), zinc-plated rivets (E ) and zinc-galvanised spikes (E ) were assigned randomly on two interconnected EIT belts. Gel was applied to the cranial or caudal belt and placed on 17 standing cattle. SCI was recorded at baseline and 3, 5, 7, 9 and 11 minutes later.
Simulations that involved electrodes with a greater skin contact area had lower and more uniform SCI. In cattle, SCI decreased with all electrodes over time (p < 0.01). Without gel, no difference was found between E and E , while SCI was higher for E (p < 0.03). With gel, SCI was lower in E and E (p < 0.026), with the SCI in E being the lowest (p < 0.01).
Low numbers of animals and static electrode position may affect SCI.
Electrode design is important for EIT measurement, with larger electrode designs able to compensate for the use of less conductive materials. Gel is not necessary to achieve acceptable SCI in large animals.
电阻抗断层成像(EIT)通过胸带式电极产生肺部通气图像。兽医需要稳健的电极设计和材料,以提供低的电极皮肤接触阻抗(SCI)。本研究旨在比较三种 EIT 电极设计和材料。
使用圆柱形、矩形和尖刺电极设计的模拟来评估电极 SCI,电极大小不一,皮肤接触不均匀。镀金垫圈(E)、镀锌铆钉(E)和镀锌刺(E)随机分配到两个互联的 EIT 带上。在颅侧或尾侧的带式电极上涂上凝胶,然后放在 17 头站立的牛身上。在基线和 3、5、7、9 和 11 分钟后记录 SCI。
模拟中具有更大皮肤接触面积的电极具有更低且更均匀的 SCI。在牛身上,随着时间的推移,所有电极的 SCI 都在下降(p < 0.01)。没有凝胶时,E 和 E 之间没有差异,而 E 的 SCI 更高(p < 0.03)。使用凝胶时,E 和 E 的 SCI 较低(p < 0.026),E 的 SCI 最低(p < 0.01)。
动物数量少和电极位置静态可能会影响 SCI。
电极设计对于 EIT 测量很重要,较大的电极设计能够补偿使用导电性较差的材料。在大型动物中,无需凝胶即可达到可接受的 SCI。