Carrera Diego A, Ricks Christian B
J Neurosurg Case Lessons. 2022 Mar 14;3(11). doi: 10.3171/CASE21731.
Cervical disc replacement (CDR) is an increasingly used alternative to fusion for symptomatic cervical disc disease. While more studies have suggested favorability of CDR over fusion procedures, limited data exist regarding implant fatigability. Here, the authors present a unique and previously unreported failure of the M6-C prosthesis causing spinal cord injury.
A 49-year-old female with history of cervical degenerative disease and prior C4-7 M6-C arthroplasty presented 9 years later after a minor fall from standing. She endorsed bilateral hand numbness ascending to forearms and shoulders, with dysesthesias and weakness. Imaging showed fractured arthroplasty penetrating the spinal cord. Revision surgery found a ruptured arthroplasty annulus with metal piece piercing the spinal cord. Partial C4 and C5 corpectomy was performed to remove the integrated fins of the arthroplasty and inspect the cord and dura. This was reconstructed with a corpectomy cage and plate. The patient made an excellent recovery, with improvement in her weakness and resolution of her sensory symptoms.
Possibility of fatigue-related failures presenting years after implantation have only been infrequently reported but can be catastrophic for patients. The authors encourage further discussions in this area, increased counseling with patients, and recommend a patient registry to better document adverse events.
对于有症状的颈椎间盘疾病,颈椎间盘置换术(CDR)作为融合术的替代方法,其应用日益广泛。虽然越来越多的研究表明CDR比融合手术更具优势,但关于植入物疲劳性的数据有限。在此,作者报告了一例独特的、此前未报道过的M6-C假体失败导致脊髓损伤的病例。
一名49岁女性,有颈椎退行性疾病病史,曾接受C4-7节段M6-C人工关节置换术,9年后在一次从站立状态轻微跌倒后就诊。她自述双侧手部麻木感向上蔓延至前臂和肩部,伴有感觉异常和无力。影像学检查显示人工关节置换术部位骨折并穿透脊髓。翻修手术发现人工关节置换术的纤维环破裂,金属碎片刺入脊髓。进行了部分C4和C5椎体次全切除,以移除人工关节置换术的融合鳍片并检查脊髓和硬脑膜。使用椎体次全切除椎间融合器和钢板进行了重建。患者恢复良好,无力症状改善,感觉症状消失。
植入多年后出现与疲劳相关的失败情况虽鲜有报道,但对患者可能是灾难性的。作者鼓励在该领域进行进一步讨论,加强对患者的咨询,并建议建立患者登记系统以更好地记录不良事件。