Increased variance in second electrode accuracy during deep brain stimulation and its relationship to pneumocephalus, brain shift, and clinical outcomes: A retrospective cohort study.

作者信息

Hart M G, Posa M, Buttery P C, Morris R C

机构信息

St George's University of London, Cranmer Terrace, London, SW17 0RE, UK.

Department of Neurosurgery, Addenbrooke's Hospital, Cambridge, CB2 0QQ, UK.

出版信息

Brain Spine. 2022 May 21;2:100893. doi: 10.1016/j.bas.2022.100893. eCollection 2022.

Abstract

•Overall electrode accuracy was 0.22+/-0.4 ​mm with only 3 (4%) electrodes out with 2 ​mm from the intended target.•Accuracy was significantly worse in the GPi versus the STN and on the second side implanted.•Inaccuracy occurred in the X (lateral) plane but was not related to pneumocephalus or brain shift.

摘要

•总体电极精度为0.22±0.4毫米,仅有3根(4%)电极偏离预定靶点2毫米以上。

•与丘脑底核相比,苍白球内侧部以及在植入的对侧,精度明显较差。

•误差出现在X(外侧)平面,但与气颅或脑移位无关。