Do newborn infants exhale through the CPAP system? Secondary analysis of a randomised cross-over trial.

机构信息

Department of Women's and Children's Health, Karolinska Institutet, Stockholm, Sweden

Department of Neonatology, Karolinska University Hospital, Stockholm, Sweden.

出版信息

Arch Dis Child Fetal Neonatal Ed. 2023 May;108(3):232-236. doi: 10.1136/archdischild-2022-324462. Epub 2022 Oct 19.

Abstract

BACKGROUND

During nasal continuous positive airway pressure (nCPAP) treatment in neonates, leakage is inevitable and can lead to reduced distending pressure in the lungs of the infant. In current practice, neither leakage nor expiratory flow is measured, which makes it difficult to assess if exhalation is through the device or entirely through leakages.

OBJECTIVE

To examine if infants treated with nCPAP exhale through the CPAP system.

DESIGN AND SETTING

Secondary data analyses from the ToNIL trial on leakages during nCPAP treatment. We retrospectively examined respiratory curves for the 50 infants included in the trial, using NI LabVIEW 2015. Each infant was measured with both prongs and nasal masks. A flow recording was classified as exhalation through the system if more than 50% of all expirations showed reverse flow, each for a minimum duration of 0.1 s.

PATIENTS

50 infants were included, born with a mean gestational age (GA) of 34 weeks, median birth weight of 1948 g and mean age at measurement 6.5 days. Inclusion criteria were CPAP treatment and a postmenstrual age (PMA) of 28-42 weeks.

RESULTS

In our measurements, 32/50 infants exhaled through the CPAP system in at least one recording with either nasal mask or prongs. Leakages exceeding 0.3 L/min were seen in 97/100 recordings.

CONCLUSIONS

During nCPAP treatment, infants can exhale through the CPAP system and leakage was common. Measuring expiratory flows and leakages in clinical settings could be valuable in optimising CPAP treatment of infants.

TRIAL REGISTRATION NUMBER

NCT03586856.

摘要

背景

在新生儿经鼻持续气道正压通气(nCPAP)治疗中,漏气流是不可避免的,这可能导致婴儿肺部的膨胀压力降低。在当前的实践中,既不测量漏气流,也不测量呼气流量,这使得很难评估婴儿是否通过设备呼气,还是完全通过漏气流呼气。

目的

检查接受 nCPAP 治疗的婴儿是否通过 CPAP 系统呼气。

设计和设置

对 nCPAP 治疗期间漏气流的 ToNIL 试验的二次数据分析。我们使用 NI LabVIEW 2015 对纳入该试验的 50 名婴儿的呼吸曲线进行了回顾性检查。每个婴儿都使用双鼻塞和鼻罩进行了测量。如果超过 50%的呼气出现反向流动,且每次持续时间至少 0.1 秒,则将流量记录分类为通过系统呼气。

患者

50 名婴儿被纳入研究,平均胎龄(GA)为 34 周,中位数出生体重为 1948 克,平均测量年龄为 6.5 天。纳入标准为 CPAP 治疗和校正胎龄(PMA)为 28-42 周。

结果

在我们的测量中,32/50 名婴儿在至少一个使用鼻罩或鼻塞的记录中通过 CPAP 系统呼气。在 100 次记录中有 97 次出现超过 0.3 升/分钟的漏气流。

结论

在 nCPAP 治疗期间,婴儿可以通过 CPAP 系统呼气,漏气流很常见。在临床环境中测量呼气流量和漏气流可能有助于优化婴儿的 CPAP 治疗。

试验注册号

NCT03586856。