在现实临床环境中使用成像光电容积脉搏波描记法对心率进行准确可靠的评估。

Accurate and Reliable Assessment of Heart Rate in Real-Life Clinical Settings Using an Imaging Photoplethysmography.

作者信息

Allado Edem, Poussel Mathias, Moussu Anthony, Hily Oriane, Temperelli Margaux, Cherifi Asma, Saunier Veronique, Bernard Yohann, Albuisson Eliane, Chenuel Bruno

机构信息

CHRU-Nancy, Exploration Fonctionnelle Respiratoire-Centre Universitaire de Médecine du Sport et Activités Physiques Adaptées, F-54000 Nancy, France.

DevAH, Université de Lorraine, F-54000 Nancy, France.

出版信息

J Clin Med. 2022 Oct 17;11(20):6101. doi: 10.3390/jcm11206101.

Abstract

Remote photoplethysmography imaging (rPPGc) is a new method measuring essential parameters, such as heart rate (HR), which uses a video camera during teleconsultation. Our work aimed to evaluate the accuracy of such remote measurements compared with existing contact point measurement methods in real-life clinical settings. The prospective hospital-based study recruited 1045 patients who required a pulmonary function test. For each patient, measurements of HR using a standard electrocardiogram acquisition system (gold standard) were carried out concomitantly with the measurements made by the rPPGc system (Caducy v1.0.0) taken within a 60 s timeframe. Age, gender, and skin phototype were collected. We performed an intraclass coefficient correlation (ICC) and Bland-Altman plotting to determine the accuracy and precision of the rPPGc algorithm readings. We achieved measurement of HR using the two methods in 963 patients. The ICC measured at a 60 s timeframe, and when we compared the rPPGc with the gold standard, it had a 95% confidence interval (CI95) value of 0.886 [0.871:0.899]. In all, 94.6% ( = 911) patients showed promising results with a CI95 in Bland-Altman plotting. Fifty-two measurements were discordant, and further analysis established the method's accuracy at 96.2%. Our results described a good accuracy and correlation between the rPPGc system and the gold standard, thus paving the way for more precise care via telemedicine.

摘要

远程光电容积脉搏波成像(rPPGc)是一种测量心率(HR)等基本参数的新方法,它在远程会诊期间使用摄像机。我们的工作旨在评估在现实临床环境中,这种远程测量与现有接触式测量方法相比的准确性。这项基于医院的前瞻性研究招募了1045名需要进行肺功能测试的患者。对于每位患者,使用标准心电图采集系统(金标准)测量心率的同时,在60秒时间内使用rPPGc系统(Caducy v1.0.0)进行测量。收集了年龄、性别和皮肤光类型。我们进行了组内相关系数(ICC)分析和布兰德-奥特曼绘图,以确定rPPGc算法读数的准确性和精确性。我们用两种方法对963名患者进行了心率测量。在60秒时间框架内测量的ICC,当我们将rPPGc与金标准进行比较时,其95%置信区间(CI95)值为0.886[0.871:0.899]。总体而言,94.6%(=911)的患者在布兰德-奥特曼绘图中显示出CI95的良好结果。52次测量结果不一致,进一步分析确定该方法的准确率为96.2%。我们的结果表明rPPGc系统与金标准之间具有良好的准确性和相关性,从而为通过远程医疗提供更精确的护理铺平了道路。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/e54e/9604655/1bdb76e654f0/jcm-11-06101-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验