Effectiveness of consumer-grade contactless vital signs monitors: a systematic review and meta-analysis.

机构信息

Department of Anesthesia and Pain Medicine, Toronto Western Hospital, University Health Network, University of Toronto, 399 Bathurst Street, McL2-405, Toronto, ON, M5T 2S8, Canada.

Institute of Medical Science, Faculty of Medicine, University of Toronto, Toronto, ON, Canada.

出版信息

J Clin Monit Comput. 2022 Feb;36(1):41-54. doi: 10.1007/s10877-021-00734-9. Epub 2021 Jul 9.

Abstract

The objective of this systematic review and meta-analysis was to analyze the effectiveness of contactless vital sign monitors that utilize a consumer-friendly camera versus medical grade instruments. A multiple database search was conducted from inception to September 2020. Inclusion criteria were as follows: studies that used a consumer-grade camera (smartphone/webcam) to examine contactless vital signs in adults; evaluated the non-contact device against a reference medical device; and used the participants' face for measurement. Twenty-six studies were included in the review of which 16 were included in Pearson's correlation and 14 studies were included in the Bland-Altman meta-analysis. Twenty-two studies measured heart rate (HR) (92%), three measured blood pressure (BP) (12%), and respiratory rate (RR) (12%). No study examined blood oxygen saturation (SpO). Most studies had a small sample size (≤ 30 participants) and were performed in a laboratory setting. Our meta-analysis found that consumer-grade contactless vital sign monitors were accurate in comparison to a medical device in measuring HR. Current contactless monitors have limitations such as motion, poor lighting, and lack of automatic face tracking. Currently available consumer-friendly contactless monitors measure HR accurately compared to standard medical devices. More studies are needed to assess the accuracy of contactless BP and RR monitors. Implementation of contactless vital sign monitors for clinical use will require validation in a larger population, in a clinical setting, and expanded to encompass other vital signs including BP, RR, and SpO.

摘要

本系统评价和荟萃分析的目的是分析使用消费者友好型摄像头的非接触式生命体征监测仪与医疗级仪器的有效性。从开始到 2020 年 9 月进行了多数据库搜索。纳入标准如下:使用消费级摄像头(智能手机/网络摄像头)检查成人非接触生命体征的研究;将非接触设备与参考医疗设备进行评估;并使用参与者的面部进行测量。共有 26 项研究纳入了综述,其中 16 项研究纳入了 Pearson 相关分析,14 项研究纳入了 Bland-Altman 荟萃分析。22 项研究测量了心率(HR)(92%),3 项研究测量了血压(BP)(12%),12%研究测量了呼吸率(RR)。没有研究检测血氧饱和度(SpO)。大多数研究样本量较小(≤30 名参与者),且在实验室环境中进行。我们的荟萃分析发现,与医疗设备相比,消费级非接触式生命体征监测仪在测量 HR 方面具有较高的准确性。当前的非接触式监测器存在一些局限性,例如运动、光照不佳和缺乏自动面部跟踪。目前可用的消费友好型非接触式监测器在测量 HR 方面与标准医疗设备相比具有较高的准确性。需要更多的研究来评估非接触式 BP 和 RR 监测器的准确性。将非接触式生命体征监测器应用于临床需要在更大的人群中、在临床环境中进行验证,并扩展到包括其他生命体征,如 BP、RR 和 SpO。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/5c54/8266631/253ae81370bf/10877_2021_734_Fig1_HTML.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验