Suppr超能文献

经 Watchman FLX 植入后的心脏计算机断层扫描:与器械相关的血栓还是器械愈合?

Cardiac computed tomography following Watchman FLX implantation: device-related thrombus or device healing?

机构信息

Department of Cardiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark.

Research and Development, Boston Scientific, Maple Grove, MN, USA.

出版信息

Eur Heart J Cardiovasc Imaging. 2023 Jan 23;24(2):250-259. doi: 10.1093/ehjci/jeac222.

Abstract

AIMS

Cardiac computed tomography (CT) is increasingly utilized during follow-up after left atrial appendage closure (LAAC). Hypoattenuated thickening (HAT) is a common finding and might represent either benign device healing or device-related thrombosis (DRT). The appearance and characteristics of HAT associated with the Watchman FLX have not been previously described. Therefore, we sought to investigate cardiac CT findings during follow-up after Watchman FLX implantation with a focus on HAT and DRT.

METHODS AND RESULTS

Retrospective single-centre, observational study including all patients with successful Watchman FLX implantation and follow-up cardiac CT between March 2019 and September 2021 (n = 244). Blinded analysis of CT images was performed describing the localization, extent, and morphology of HAT and correlated to imaging and histology findings in a canine model. Relevant clinical and preclinical ethical approvals were obtained.Overall, HAT was present in 156 cases (64%) and could be classified as either subfabric hypoattenuation (n = 59), flat sessile HAT (n = 78), protruding sessile HAT (n = 16), or pedunculated HAT (n = 3). All cases of pedunculated HAT and five cases of protruding sessile HAT were considered as high-grade HAT (n = 7). Subfabric hypoattenuation and flat sessile HAT correlated with device healing and endothelialization in histological analysis of explanted devices.

CONCLUSION

Subfabric hypoattenuation and flat sessile HAT are frequent CT findings for Watchman FLX, likely representing benign device healing and endothelialization. Pedunculated HAT and protruding HAT are infrequent CT findings that might represent DRT.

摘要

目的

左心耳封堵(LAAC)术后的随访中越来越多地应用心脏计算机断层扫描(CT)。低衰减增厚(HAT)是一种常见的表现,可能代表良性的器械愈合或器械相关血栓形成(DRT)。以前尚未描述与 Watchman FLX 相关的 HAT 的外观和特征。因此,我们试图研究 Watchman FLX 植入后的随访中心脏 CT 的发现,重点是 HAT 和 DRT。

方法和结果

这是一项回顾性的单中心观察性研究,纳入了 2019 年 3 月至 2021 年 9 月间所有成功植入 Watchman FLX 并接受随访心脏 CT 的患者(n=244)。对 CT 图像进行盲法分析,描述 HAT 的定位、程度和形态,并与犬模型中的影像学和组织学发现相关联。获得了相关的临床和临床前伦理批准。总的来说,156 例(64%)存在 HAT,可分为亚结构低衰减(n=59)、扁平贴壁性 HAT(n=78)、突出贴壁性 HAT(n=16)或带蒂 HAT(n=3)。所有带蒂 HAT 和 5 例突出贴壁性 HAT 均被认为是高级 HAT(n=7)。亚结构低衰减和扁平贴壁性 HAT 在对取出的器械进行组织学分析时与器械愈合和内皮化相关。

结论

亚结构低衰减和扁平贴壁性 HAT 是 Watchman FLX 的常见 CT 表现,可能代表良性的器械愈合和内皮化。带蒂 HAT 和突出 HAT 是不常见的 CT 表现,可能代表 DRT。