一种新设计的角度可调胶片夹在减少牙科学学生水平管移位技术中的锥形切割误差和节省时间方面的效果。

Efficacy of a newly designed angulation-adjustable film holder for reducing cone-cutting errors and saving time in horizontal tube-shift technique by dental students.

作者信息

Puapichartdumrong Peraya, Eakpunyakul Nongnapat, Tanpumiprathet Suphakarn, Khueankaew Pimrekha, Saelim Priawwan, Piyapattamin Thosapol

机构信息

Faculty of Dentistry, Naresuan University, Phitsanulok 65000, Thailand.

出版信息

Heliyon. 2022 Nov 12;8(11):e11567. doi: 10.1016/j.heliyon.2022.e11567. eCollection 2022 Nov.

Abstract

OBJECTIVES

To compare the cone-cutting error frequency and the X-ray cone alignment time in the horizontal tube-shift technique between the newly designed angulation-adjustable and the extension cone paralleling (XCP) holders, by dental students.

MATERIALS AND METHODS

Two film holders were assigned for a random test. The mandibular left first molar position of a laboratory phantom head was used. Intraoral periapical radiography was performed horizontally at right-angle, 20° mesial-, and 20° distal projections by fifth-year dental students (n = 41). The cone-cutting error frequency and the X-ray cone alignment time were measured and analyzed statistically at a significant level of < 0.05.

RESULTS

Using the two holders at right angle caused no significant difference ( > 0.05) in the cone-cutting error frequency or the X-ray cone alignment time. At the horizontal tube-shift angles, some significantly greater frequencies of cone-cutting errors at both 20° mesial (43.9%) and 20° distal (73.2%) shifts were detected in the XCP group, but none in the angulation-adjustable group. For X-ray cone alignment time at both 20° mesial and 20° distal shifts, the XCP group spent a significantly longer time ( < 0.05) than the angulation-adjustable group.

CONCLUSIONS

The usages of the angulation-adjustable holder in the horizontal tube-shift technique resulted in none of the cone-cutting error and a significant reduction of time for the X-ray cone alignment, when compared to those of the XCP instrument.

CLINICAL SIGNIFICANCE

The angulation-adjustable holder effectively reduced cone cutting error and treatment time, both of which were beneficial to the patients and the dental personnel.

摘要

目的

比较牙科学生使用新设计的角度可调式持片器和延伸锥平行(XCP)持片器时,水平管移位技术中的锥形切割误差频率和X射线锥对准时间。

材料与方法

随机分配两种持片器进行测试。使用实验室模型头部的下颌左侧第一磨牙位置。由五年级牙科学生(n = 41)进行口腔内根尖周X线摄影,分别在直角、近中20°和远中20°投影下水平进行。测量并统计分析锥形切割误差频率和X射线锥对准时间,显著性水平为<0.05。

结果

在直角使用两种持片器时,锥形切割误差频率或X射线锥对准时间均无显著差异(>0.05)。在水平管移位角度时,XCP组在近中20°(43.9%)和远中20°(73.2%)移位时的锥形切割误差频率均显著更高,但角度可调组均未出现。对于近中20°和远中20°移位时的X射线锥对准时间,XCP组花费的时间显著长于角度可调组(<0.05)。

结论

与XCP器械相比,在水平管移位技术中使用角度可调式持片器不会产生锥形切割误差,且显著减少了X射线锥对准时间。

临床意义

角度可调式持片器有效减少了锥形切割误差和治疗时间,这对患者和牙科工作人员均有益。