To compare the cone-cutting error frequency and the X-ray cone alignment time in the horizontal tube-shift technique between the newly designed angulation-adjustable and the extension cone paralleling (XCP) holders, by dental students.
Two film holders were assigned for a random test. The mandibular left first molar position of a laboratory phantom head was used. Intraoral periapical radiography was performed horizontally at right-angle, 20° mesial-, and 20° distal projections by fifth-year dental students (n = 41). The cone-cutting error frequency and the X-ray cone alignment time were measured and analyzed statistically at a significant level of < 0.05.
Using the two holders at right angle caused no significant difference ( > 0.05) in the cone-cutting error frequency or the X-ray cone alignment time. At the horizontal tube-shift angles, some significantly greater frequencies of cone-cutting errors at both 20° mesial (43.9%) and 20° distal (73.2%) shifts were detected in the XCP group, but none in the angulation-adjustable group. For X-ray cone alignment time at both 20° mesial and 20° distal shifts, the XCP group spent a significantly longer time ( < 0.05) than the angulation-adjustable group.
The usages of the angulation-adjustable holder in the horizontal tube-shift technique resulted in none of the cone-cutting error and a significant reduction of time for the X-ray cone alignment, when compared to those of the XCP instrument.
The angulation-adjustable holder effectively reduced cone cutting error and treatment time, both of which were beneficial to the patients and the dental personnel.
比较牙科学生使用新设计的角度可调式持片器和延伸锥平行(XCP)持片器时,水平管移位技术中的锥形切割误差频率和X射线锥对准时间。
随机分配两种持片器进行测试。使用实验室模型头部的下颌左侧第一磨牙位置。由五年级牙科学生(n = 41)进行口腔内根尖周X线摄影,分别在直角、近中20°和远中20°投影下水平进行。测量并统计分析锥形切割误差频率和X射线锥对准时间,显著性水平为<0.05。
在直角使用两种持片器时,锥形切割误差频率或X射线锥对准时间均无显著差异(>0.05)。在水平管移位角度时,XCP组在近中20°(43.9%)和远中20°(73.2%)移位时的锥形切割误差频率均显著更高,但角度可调组均未出现。对于近中20°和远中20°移位时的X射线锥对准时间,XCP组花费的时间显著长于角度可调组(<0.05)。
与XCP器械相比,在水平管移位技术中使用角度可调式持片器不会产生锥形切割误差,且显著减少了X射线锥对准时间。
角度可调式持片器有效减少了锥形切割误差和治疗时间,这对患者和牙科工作人员均有益。