Method Verification of the Caretium XC-A30 Automated Erythrocyte Sedimentation Rate Analyser for Erythrocyte Sedimentation Rate.

作者信息

Khongwichit Soemwit, Saelim Menapha, Na-Songkhla Yanisa, Buncherd Hansuk, Nopparatana Chawadee, Srinoun Kanitta

机构信息

Faculty of Medical Technology, Prince of Songkla University, Songkhla, Thailand.

Department of Pathology, Faculty of Medicine, Prince of Songkla University, Songkhla, Thailand.

出版信息

Malays J Med Sci. 2022 Oct;29(5):39-47. doi: 10.21315/mjms2022.29.5.5. Epub 2022 Oct 28.

Abstract

BACKGROUND

The erythrocyte sedimentation rate (ESR) analyser is widely used in haematological testing. In addition to the Westergren method, new automatic methods for ESR measurements have been developed. We aimed to study the reliability, precision, accuracy and stability of the Caretium XC-A30 automated ESR analyser.

METHODS

Ethylenediamine tetraacetic acid (EDTA)-treated blood samples were analysed via the Caretium XC-A30 automated ESR analyser and the Westergren method to compare accuracy. Precision was assessed using control samples and patient samples were classified into three groups-low, medium and high-according to their rates of sedimentation. Moreover, a stability test was performed.

RESULTS

The correlation coefficient of the results of the Caretium XC-A30 and Westergren analyses was 0.97. The correlation coefficient of ESR values obtained from the two methods assessed in the low, medium and high groups were r = 0.80, = 0.68 and = 0.74, respectively. The coefficient of variation of within-run (%CVw) and between-run (%CVb), with replicates performed with commercial controls samples, were 7.54% and 8.04% for the normal control and 4.68% and 3.50% for abnormal control, respectively. The %CVw obtained with patient samples in the low, medium and high groups were 10.68%, 13.13% and 4.45%, respectively. The Caretium XC-A30 measurements were stable for up to 24 h when samples were stored at 4 °C.

CONCLUSION

The Caretium XC-A30 ESR analyser proved to be a suitable instrument for routine analysis of ESR.

摘要

背景

红细胞沉降率(ESR)分析仪在血液学检测中广泛应用。除魏氏法外,已开发出用于ESR测量的新的自动方法。我们旨在研究Caretium XC - A30自动ESR分析仪的可靠性、精密度、准确性和稳定性。

方法

通过Caretium XC - A30自动ESR分析仪和魏氏法对乙二胺四乙酸(EDTA)处理的血样进行分析以比较准确性。使用对照样品评估精密度,根据沉降率将患者样品分为低、中、高三组。此外,进行了稳定性测试。

结果

Caretium XC - A30分析结果与魏氏法分析结果的相关系数为0.97。在低、中、高组中,两种方法获得的ESR值的相关系数分别为r = 0.80、r = 0.68和r = 0.74。使用商业对照样品进行重复测量时,批内变异系数(%CVw)和批间变异系数(%CVb),正常对照分别为7.54%和8.04%,异常对照分别为4.68%和3.50%。低、中、高组患者样品的%CVw分别为10.68%、13.13%和4.45%。当样品在4℃储存时,Caretium XC - A30测量在长达24小时内是稳定的。

结论

Caretium XC - A30 ESR分析仪被证明是用于ESR常规分析的合适仪器。