Research and Development, Science & Emerging Technologies, Aerogen Limited, Galway Business Park, Galway, Ireland.
Neonatology Department, Guangdong Women and Children Hospital, Guangdong Neonatal ICU Medical Quality Control Center, Guangdong, China.
Pediatr Pulmonol. 2023 Mar;58(3):878-886. doi: 10.1002/ppul.26270. Epub 2022 Dec 12.
Aerosol drug delivery during high flow nasal oxygen (HFNO) and invasive mechanical ventilation (IMV) are key respiratory care strategies available for the treatment of pediatric patients. We aimed to quantify the impact of different HFNO and IMV set-ups on tracheal drug delivery via a vibrating mesh nebuliser (VMN).
Percent tracheal dose via VMN was quantified during HFNO therapy and IMV in a benchtop model of a 9-month-old infant. Under HFNO, 3 cannula sizes were used at 3 flow rate settings with the VMN placed at the dry side of the humidifier. Under IMV, tracheal dose when VMN was placed at the dry side of the humidifier, 15 cm from the wye and between the wye and endotracheal tube (ETT) was assessed. Salbutamol at 2.5 mg/2.5 ml (1 mg/ml) was used for each test (N = 5). The impact of VMN refill on circuit pressure under HFNO and IMV was also assessed.
Tracheal dose was highest during HFNO with the largest cannula size (OPT318) set to the lowest flow rate setting of 2 L/min (liter per minute) (5.80 ± 0.17%). Increasing flow rate reduced tracheal drug delivery for all cannulas. For IMV, VMN on the dry side of the humidifier and between the wye and ETT gave optimal drug delivery (4.49 ± 0.14% vs. 4.43 ± 0.26% respectively). VMN refill did not impact circuit pressure for either HFNO therapy or IMV.
Gas flow rate and cannula size during HFNO and VMN position during IMV has a significant effect on tracheal drug delivery in a pediatric setting.
在高流量鼻氧(HFNO)和有创机械通气(IMV)期间进行气溶胶药物输送是治疗儿科患者的关键呼吸治疗策略。我们旨在通过振动网孔雾化器(VMN)量化不同 HFNO 和 IMV 设置对气管药物输送的影响。
在 9 个月大婴儿的台式模型中,在 HFNO 治疗和 IMV 期间通过 VMN 量化气管内药物输送的百分比。在 HFNO 下,在 3 个流量设置下使用 3 个套管大小,VMN 放置在加湿器的干燥侧。在 IMV 下,评估 VMN 放置在加湿器的干燥侧、距 Y 型管 15 厘米和 Y 型管和气管内导管(ETT)之间时的气管剂量。每次测试均使用沙丁胺醇 2.5 毫克/2.5 毫升(1 毫克/毫升)(N=5)。还评估了 VMN 再填充对 HFNO 和 IMV 下回路压力的影响。
在 HFNO 下,最大套管尺寸(OPT318)设置为最低流量设置 2 L/min(升/分钟)时,气管内药物输送量最高(5.80±0.17%)。对于所有套管,增加流量会降低气管内药物输送。对于 IMV,加湿器干燥侧和 Y 型管与 ETT 之间的 VMN 提供了最佳的药物输送(分别为 4.49±0.14%和 4.43±0.26%)。VMN 再填充对 HFNO 治疗或 IMV 下的回路压力均无影响。
HFNO 期间的气体流速和套管尺寸以及 IMV 期间 VMN 的位置对儿科环境下的气管药物输送有显著影响。